The Agilent/HP E6000B Mainframe maximizes your network uptime by locating and characterizing faults quickly and accurately.

The combination of 16,000 data points and a minimum sample spacing of 8cm allows the powerful analysis algorithm to determine the exact location and characteristic of an event.

The Agilent/HP E6000B OTDR repeatedly outputs an optical pulse into the connected fiber and measures the reflections from this pulse. The trace displayed on the screen is a graph of this reflected power (backscatter) as a function of the distance along the fiber.

Events are changes in the fiber causing the trace to deviate from a straight line. Events can be Reflective or Non-Reflective.

Reflective Events occur when some of the pulse energy is reflected, for example at a connector. Non-Reflective Events occur at parts of the fiber where there is some loss but no light is reflected. Non-Reflective Events produce a dip on the trace.

The E6000B OTDR calculates the distance of such an "event" in the fiber from the time it takes the reflected signal to return. The further away an event is, the longer it takes for its reflection to return to the E6000B OTDR.

A fully automatic function is available and the E6000B OTDR sets itself up to achieve the best results for the measurements such as the location (distance) of events on the link, the end of the link or a break, the attenuation coefficient of the fiber in the link, the loss of an individual event or the total end-to-end loss of the link, the magnitude of the reflection (or reflectance) of an event, such as a connector, the cumulative loss to an event can be measured automatically.

The Agilent E6000B can operate through the supplied AC adapter between 100V and 240V ±10%, at a frequency in the range from 50 to 60 Hz. The maximum power consumption is 30VA with all options installed.

The Agilent E6000B can operate from a DC power source that supplies between 16V and 24V. The maximum power consumption during a quick charge is 30W with all options installed. Typical power consumption is below 8W.

Key Points:

  • Built in Applications
  • Built in multi fiber test
  • With template mode
  • One button operation
  • Automatic event adjustment
  • 15 secs power up cycle
  • Automatic fiber characteristic

Brand Agilent / HP
Model E6000B
Description Mini OTDR Mainframe
Category OTDR
Sub Category N/A
Specifications:
Form Factor Module
Fiber Type Multimode
Wavelength Range 850nm to 1300nm
Maximum Dynamic Range 18dB
User Interface Proprietary
Input Power Universal
Inter memory Up to 200 traces
Operating Temperature 0 to 50 deg C
Storage Temperature -40 to 60 deg C
Dimensions 194 mm H, 290 mm W, 75 mm D (7.7" x 11.4" x 3.0")
Weight net < 2.9 kg (6.4 lbs)

Agilent E6005A
Central Wavelength 850±30 nm/ 1300±30 nm
Applicable Fiber multimode 62.5 µm
Pulsewidth 10 ns 100 ns 1µs 10 µs
Dynamic Range 9/17 26/22 - /28 - /34
Event Deadzone 3m
Attenuation Deadzone 10m
Agilent HP E6000B Mini-OTDR With E6005A Multimode 850nm & 1300nm Module

Includes An EXFO FLS-235B Pocket VFL ($500 Value), Carrying Case & Accessories As Listed!

Agilent E6000B E6005A E-6000B E-6005A OTDR Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.
  • Agilent E6000B OTDR Mainfame
  • Agilent E6005A MultiMode Module
    • 850nm
    • 1300nm
  • EXFO FLS-235B Pocket VFL
  • Fiber Cables
    • (2) MM ST-SC
    • (2) MM ST-ST
  • Battery Pack
  • AC Adapter/Charger
  • Power Cord
  • RS-232 PC Interface Cable
  • User Manual On CD-ROM
  • Soft Carrying Case