The Agilent/HP E6000B Mainframe maximizes your network uptime by locating and
characterizing faults quickly and accurately.
The combination of 16,000 data points and a minimum sample spacing of 8cm
allows the powerful analysis algorithm to determine the exact location and
characteristic of an event.
The Agilent/HP E6000B OTDR repeatedly outputs an optical pulse into the
connected fiber and measures the reflections from this pulse. The trace
displayed on the screen is a graph of this reflected power (backscatter) as a
function of the distance along the fiber.
Events are changes in the fiber causing the trace to deviate from a straight
line. Events can be Reflective or Non-Reflective.
Reflective Events occur when some of the pulse energy is reflected, for
example at a connector. Non-Reflective Events occur at parts of the fiber where
there is some loss but no light is reflected. Non-Reflective Events produce a
dip on the trace.
The E6000B OTDR calculates the distance of such an "event" in the fiber from
the time it takes the reflected signal to return. The further away an event is,
the longer it takes for its reflection to return to the E6000B OTDR.
A fully automatic function is available and the E6000B OTDR sets itself up to
achieve the best results for the measurements such as the location (distance) of
events on the link, the end of the link or a break, the attenuation coefficient
of the fiber in the link, the loss of an individual event or the total
end-to-end loss of the link, the magnitude of the reflection (or reflectance) of
an event, such as a connector, the cumulative loss to an event can be measured
automatically.
The Agilent E6000B can operate through the supplied AC adapter between 100V
and 240V ±10%, at a frequency in the range from 50 to 60 Hz. The maximum power
consumption is 30VA with all options installed.
The Agilent E6000B can operate from a DC power source that supplies between
16V and 24V. The maximum power consumption during a quick charge is 30W with all
options installed. Typical power consumption is below 8W.
Key Points:
- Built in Applications
- Built in multi fiber test
- With template mode
- One button operation
- Automatic event adjustment
- 15 secs power up cycle
- Automatic fiber characteristic
| Brand |
Agilent / HP |
| Model |
E6000B |
| Description |
Mini OTDR Mainframe |
| Category |
OTDR |
| Sub Category |
N/A |
| Specifications: |
| User Interface |
Proprietary |
| Input Power |
Universal |
| Inter memory |
Up to 200 traces |
| Operating Temperature |
0 to 50 deg C |
| Storage Temperature |
-40 to 60 deg C |
| Dimensions |
194 mm H, 290 mm W, 75 mm D (7.7" x 11.4" x 3.0") |
| Weight |
net < 2.9 kg (6.4 lbs) |
Agilent E6008B
| Central Wavelength |
1310 nm ± 25 nm/1550 nm ± 5 nm |
| Applicable Fiber |
Singlemode |
| Pulsewidth |
10 ns |
100 ns |
1µs |
10 µs |
20 µs |
| Dynamic Range |
24/22 |
29/27 |
35/34 |
42/41 |
45/43 |
| Event Deadzone |
3m |
| Attenuation Deadzone |
10/12m |
Agilent E6006A Power Meter
Submodule Characteristics Sensor element: InGaAs Wavelength
range: 800 ? 1650 nm Calibrated wavelengths: 850 nm, 1300 nm, 1310 nm,
1550 nm, 1625 nm (special wavelengths on request). Power range: +10 to -70
dBm Max. input power (damage level): +13 dBm/20 mW Display Resolution:
0.01 dB Display Units: dBm, dB, mW, µW, nW, pW Display Contents •
Calibrated ? in nm • Modulation frequency in Hz • Reference value in
dB Display Updates per second: 3 Optical input:
User-exchangeable Connector Interface Applicable fiber type: 9/125 µm,
50/125 µm, 62.5/125 µm Specifications Uncertainty at reference
conditions: ± 3% Power level: - 20 dBm Continuous Wave
(CW) Wavelength: 1300±3 nm, 1310±3 nm, 1550±3 nm Fiber type: 50/125 µm
graded index, Agilent/HMS-10 connector Spectral bandwidth: up to 10
nm Ambient temperature: +18 to + 28 °C At day of calibration (add 0.3% for
aging of over one year; add 0.6% for aging of over two years). Total
uncertainty: ± 5% ± 0.5 nW (1310, 1550 nm) Power level: +0 to ?50
dBm Continuous Wave (CW) Wavelength: 850±3 nm, 1300±3 nm, 1310±3 nm,
1550±3 nm, Fiber type: SM to 50µm graded index (add 2% to total
uncertainty for fiber 62.5µm). Straight and angled connectors Ambient
temperature: +10 to +40°C Within 2 years after calibration Supplementary
Performance Characteristics • Automatic Zeroing Circuitry • Automatic
Ranging • Modulation frequency recognition (270 Hz, 1 kHz, 2 kHz) is
available at power level between +10 and ?45 dBm (peak amplitude). •
Dual Wavelength measurement is available at power levels between +10 and
?45 dBm (peak amplitude). • Wavelength encoding recognition (350 Hz,
550 Hz) is available at power levels between +10 and ?45 dBm (peak
amplitude). • Reference value is presettable from +30 to ?80 dBm • Each
calibrated wavelength has its own reference memory. • The actual display
content can be transferred to reference memory(DISP?REF). • Hold Data
functionality General Specifications Dimensions: ca. 120 mm H x 40 mm
W x 25 mm D (4.7? x 1.6? x 1.0?) Weight: <130g. Operating
Temperature: 0 to +50 °C Storage Temperature: -40 to +60 °C Humidity: 95%
R.H. from 0°C to 40°C non cond.
|