Fluke ESA612 kit with accessories

Fluke ESA612 Electrical Safety Analyzer 115V ac

Fluke ESA615 kit with accessories

Fluke ESA615 115V ac Electrical Safety Analyzer

Fluke ESA612 Electrical Safety Analyzer 230V ac

$3,495.00
Availability: Out of stock
SKU
MT08

Condition: Refurbished

Warranty: 1 YEAR WARRANTY

Request a Quote for

Have Questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

we are commited to win your business!

 Fluke Networks ESA612 Electrical Safety Analyzer
The ESA612 Electrical Safety Analyzer represents the next generation in testers for biomedical professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.

With selection of two test loads, this versatile product can be used worldwide to test to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751), and AN/NZS 3551.

The versatility of the multifaceted ESA612 is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.


Features:
  • Portable, ergonomic, lightweight and easy to use
  • Large, easy-to-read display with adjustable contrast
  • Human-factors-designed user interface
  • Bail design for stand-up testing in field environments
  • Five applied parts jacks and easy ECG snap connection with optional expander box
  • ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool
  • Replaceable mains fuses keep the device in the field and out of the repair shop
  • Internal memory for 100 test records
  • USB connection for use with Ansur and Data Viewer software (for memory download to PC)
  • Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
  • Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval

Specifications:
Voltage
Range (mains voltage)  180 to 264 V ac rms
Range (accessible voltage) 0 V ac to 300 V ac rms
Accuracy ± (2 % of reading + 0.2 V)
Voltage tests Mains, accessible, and point-to-point
Earth resistance
Mode Two terminal
Test current > 200 mA ac
Range 0 W to 2 W
Accuracy ± (2 % of reading + 0.015 W)
Resistance tests Earth resistance and point-to-point
Equipment current
Mode AC rms
Range 0 A to 20 A
Accuracy ± 5 % of reading + (2 counts or 0.2 A, whichever is greater)
Duty cycle 15 A to 20 A, 5 min on/5 min off
10 A to 15 A, 7 min on/3 min off
0 A to 10 A continuous
Leakage current
Modes
Modes are available in all leakage tests with the exception
of MAP leakages that are available only in true-rms
AC + DC (true-rms)
AC only
DC only
Patient load selection
(input impedance)
AAMI ES1-1993
IEC 60601
Crest factor = 3
Ranges 0 µA to 199.9 µA
200 µA to 1999 µA
2 mA to 10 mA
Frequency response/accuracy
DC to 1 kHz
1 kHz to 100 kHz
100 kHz to 1 MHz

± (1 % of reading + (1 µA or 1 LSD, whichever is greater))
± (2 % of reading + (1 µA or 1 LSD, whichever is greater))
± (5 % of reading + (1 µA or 1 LSD, whichever is greater))
Leakage tests Ground wire (earth)
Chassis (enclosure)
Lead to ground (patient)
Lead to lead (patient auxiliary)
Lead isolation (mains on applied part)
Direct equipment
Direct applied part
Alternative equipment
Alternative applied part
Point to point
Mains on applied part test voltage 100 % of mains
Differential leakage
Ranges 10 µA to 199 µA
200 µA to 2000 µA
2 mA to 20 mA
Accuracy ± 10 % of reading + (2 counts or 20 µA, whichever is greater)
Insulation resistance
Ranges 0.5 MW to 20 MW
20 MW to 100 MW
Accuracy ± (2 % of reading + 0.2 MW)
± (7.5 % of reading + 0.2 MW)
Source test voltage 500 V dc
250 V dc
Insulation resistance tests Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible
conductive part) and AP- NE (non-earthed accessible conductive
part)
ECG performance waveforms
Accuracy ± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV
Lead II configuration
Waveforms
ECG complex (BPM)
Ventricular fibrillation
Square wave (50 % duty cycle) (Hz)
Sine wave (Hz)
Triangle wave (Hz)
Pulse (63 ms pulse width)

30, 60, 120, 180, and 240

0.125 and 2
10, 40, 50, 60, and 100
2
30 BPM and 60 BPM
Power ratings
Mains voltage outlet 230 V ac
Mains voltage inlet power range 180 to 264 V ac rms
Maximum current 16 A
Hz 50 or 60
Environmental specifications
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity 10 % to 90 % non-condensing
Altitude 230 V ac mains supply voltage up to 2,000 m
 




For full Fluke ESA612 product specifications, please click here: Fluke ESA612

Fluke ESA612 230V ac Electrical Safety Analyzer

The ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool

With selection of two test loads, this versatile product can be used worldwide to test to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751), and AN/NZS 3551

Includes Carrying Case & Accessories As Listed

ESA 612 ESA-612

Fluke ESA612 ESA 612 Electrical Safety Analyzer Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.

 

BC# 46006-L/T*

  • Fluke ESA612 European 230V ac Electrical Safety Analyzer (BC# 46006-L/T*)  
  • (2) Hard Point Test Leads
  • USB PC Interface Cable
  • Power Cord
  • User Manual On CD-ROM
  • Carrying Case (G)