HP 16192A Parallel Electrode SMD Test Fixture
$0.00
Availability:
Out of stock
SKU
KE04
Condition: Refurbished
Warranty: 90 Day Exchange
Shipping: USA & Worldwide
Request a Quote for
- HP 16192A Electrode SMD Test Fixture (BC# 21390)
- Magnifying Glass
HP 16192A Parallel Electrode SMD Test Fixture
The 16192A test fixure is designed for impedance evaluations of side electrode SMD components. The minimum SMD size that this fixure is adapted to evaluate is 1(L)[mm]
Compensation and Measurement
Open and short compensations are recommended in combination with the electrical length compensation before measurement. The fixtureメs electrical length must be entered into the electrical length compensation function of the measurement instrument first. Then open compensation is performed by separating the high and the low electrodes from each other. The separation should be equivalent in size to the DUTメs width. Short compensation is performed by using option 16192A-010/701 shorting bar set. After performing open and short compensations in combination with the electrical length compensation, the DUT is inserted into the test fixture.
Specifications:
For full Agilent 16192A specifications, please click here: Agilent 16192A
The 16192A test fixure is designed for impedance evaluations of side electrode SMD components. The minimum SMD size that this fixure is adapted to evaluate is 1(L)[mm]
Compensation and Measurement
Open and short compensations are recommended in combination with the electrical length compensation before measurement. The fixtureメs electrical length must be entered into the electrical length compensation function of the measurement instrument first. Then open compensation is performed by separating the high and the low electrodes from each other. The separation should be equivalent in size to the DUTメs width. Short compensation is performed by using option 16192A-010/701 shorting bar set. After performing open and short compensations in combination with the electrical length compensation, the DUT is inserted into the test fixture.
Specifications:
Terminal Connector: | 7 mm |
DUT Connection: | 2-Terminal |
Electrical Length: | 11 mm |
Dimensions (approx.) | 150(W) x 70(H) x 90(D) [mm] |
Weight (approx.): | 400 g |
Proportional Error | 1.5 x f2 [%] |
Open Repeatability | 2 + 30 x f [µS] |
Short Repeatability | 30 + 250 x f [mO] |
Applicable Instrument: | 4291B*, 4294A + 42942A, (4287A, E4982A, E5061B-3L5 with Opt. E50616B-005 + 16201A, E4991A applicable in a limited frequency range.) When used with 16085B: 4263B, 4268A, 4279A, 4284A, 4285A, 4288A, E4980A, E4981A |
Frequency | DC to 2 GHz |
Maximum Voltage: | ±42 V peak max. (AC+DC) |
Operating Temperature | -55°C to +85°C |
DUT Size | 1 mm to 20 mm (length) |
For full Agilent 16192A specifications, please click here: Agilent 16192A