Agilent 33250A Function/Arbitrary Waveform Generator
Standard Waveforms
The Agilent Technologies 33250A Function/Arbitrary Waveform Generator uses direct digital-synthesis techniques to create a stable, accurate output on all waveforms, down to 1 µHz frequency resolution. The benefits are apparent in every signal you produce, from the sine wave frequency accuracy to the fast rise/fall times of square waves, to the ramp linearity.
Front-panel operation of the 33250A is straightforward and user friendly. The knob or numeric keypad can be used to adjust frequency, amplitude and offset. You can even enter voltage values directly in Vpp, Vrms, dBm, or high/low levels. Timing parameters can be entered in hertz (Hz) or seconds.
Custom Waveform Generation
Why settle for a basic function generator when you can get arbitrary waveforms at no extra cost? With the 33250A, you can generate arbitrary waveforms with 12-bit vertical resolution, 64K memory depth, and a sample rate of 200 MSa/s. You can also store up to four 64K-deep arbitrary waveforms in non-volatile memory with user-defined names to help you find the right waveform when you need it most
The included Agilent IntuiLink software allows you to easily create, edit, and download complex waveforms using the intuiLink Arbitrary Waveform Editor. Or you can capture a waveform using IntuiLink oscilloscope or DMM and send it to the 33250A for output. For programmers, ActiveX components can be used to control the instrument using SCPI commands. IntuiLink provides the tools to easily create, download, and manage waveforms for your 33250A. To find out more about IntuiLink, visit www.agilent.com/find/intuilink.
Pulse Generation
The 33250A can generate simple pulses up to 50 MHz. With variable edge time, pulse width and voltage level, the 33250A is ideally suited to a wide variety of pulse applications
Built-in Versatility
AM, FM and FSK capabilities make it easy to modulate waveforms with or without a separate source. Linear or logarithmic sweeps can be performed with a programmable frequency marker signal. Programmable burst count and gating allow you to further customize your signal.
For system applications, both GPIB and RS-232 interfaces are standard, and support full programmability using SCPI commands.
Color Graphical Display
The unique design of the 33250A combines a low-profile instrument with the benefits of a color graphical display. Now you can display multiple waveform parameters at the same time. The graphical interface also allows you to modify arbitrary waveforms quickly and easily.
Timebase Stability and Clock Reference
The 33250A TCXO timebase gives you frequency accuracy of 2 ppm for your most demanding applications. The external clock reference input/output lets you synchronize to an external 10 MHz clock, to another 33250A, or to an Agilent 33120A. Phase adjustments can be made from the front panel or via a computer interface, allowing precise phase calibration and adjustment.
Features:
- 80 MHz sine and square wave outputs
- Sine, square, ramp, noise and other waveforms
- 50 MHz pulse waveforms with variable rise/fall times
- 12-bit, 200 MSa/s, 64K-point deep arbitrary waveform
Specifications:
WAVEFORMS | |
Standard | sine, square, pulse, ramp, noise, sin(x)/x, exponential rise, exponential fall, cardiac, DC volts |
Arbritrary | |
Waveform length | 1 to 64K points |
Amplitude resolution | 12 bits (including sign) |
Repetition rate | 1 µHz to 25 MHz |
Sample rate | 200 MSa/s |
Filter bandwidth | 50 MHz |
Non-vol. Memory | Four (4) 64K waveforms |
FREQUENCY CHARACTERISTICS | |
Sine | 1 µHz to 80 MHz |
Square | 1 µHz to 80 MHz |
Pulse | 500 µHz to 50 MHz |
Arb | 1 µHz to 25 MHz |
Ramp | 1 µHz to 1 MHz |
White noise | 50 MHz bandwidth |
Resolution | 1 µHz; except pulse, 5 digits |
Accuracy (1 Year) | 2 ppm, 18°C to 28°C 3 ppm, 0°C to 55°C |
GENERAL | |
Power supply | 100-240 V, 50-60 Hz 100-127 V, 50-400 Hz |
Power soncumption | 140 VA |
Operating temperature | 0°C to 55°C |
Storage temperature | -30°C to 70°C |
Stored states | 4 named user configurations |
Power on state | default or last |
Interface | IEEE-488 and RS-232 std |
Language | SCPI-1997, IEEE-488.2 |
Dimensions (WxHxD) | Bench top: 254 x 104 x 374 mm Rackmount: 213 x 89 x 348 mm |
Weight | 4.6 kg |
Safety designed to | EN61010-1, CSA1010.1, UL-311-1 |
EMC Tested to | IEC-61326-1 IEC-61000-4-3 criteria B IEC-61000-4-6 criteria B |
Vibration and shock | MIL-T-28800E, Type III, Class 5 |
Acoustic noise | 40 dBA |
Warm-up time | 1 Hour |
Calibration interval | 1 Year |