- Allied Healthcare Autovent 4000 L760 Transport Ventilator (BC# -L/T*)
Request a Quote for
Allied Healthcare Autovent 4000 Transport Ventilator
- Pnumatically powered.. durable and light, with all basic models weighing less than 5lbs.
- Independent settings for the critical ventilation parameters of Breaths Per Minute, Inspiratory Time, Tidal Volume, and Airway Pressure Relief.
- Manual Breath Button and Anti-Breath Stacking standard on all models
- Large color-coded Manometer Gauge along with a full set of alarms to ensure optimal performance and patient safety
L760 operations
- Pneumatic Operation
- Adjustable Tidal Volume
- Adjustable BPM
- Child and Adult settings
- Internal Demand Valve
- Anti-Breath Stacking
- Manometer Readout
- Adjustable Pressure Relief with Alarm
- Low Source Gas Pressure Alarm
- Color-coded Labels and Gauge for Initial Settings
- Manual Breath Button
Specifications:
Gas Supply Pressure | 40.6 to 87.0 psi oxygen DISS |
Breaths per minute | Accuracy: ±10% Range: 1 second inspiratory time = 0 to 30 bpm Range: 2 second inspiratory time = 0 to 20 bpm |
Tidal Volume | Accuracy: ±10% with 100% oxygen Range: 1 second inspiratory time = 200 ml to 600 ml Range: 2 second inspiratory time = 400 ml to 1200 ml |
Inspiratory Time | Accuracy: ±10% |
Safety Pressure Relief | Adjustable from 20 to 80 cmH20 ±10% |
Low Source Gas Alarm | Activate at 275 to 262 kPa (40 to 38 psi) source pressure |
Oxygen Inlet Filter | 65 micron sintered bronze |
Burst Pressure | 250 psix minimum (1724 kPa) |
Leakage | Designed so that oxygen is not allowed to leak through any seals or fittings |
Gauge | 0-100 cm H20 (accuracy ±2% of reading) |
Inspiratory and Expiratory Resistance | 5 cm H20 maximum |
Inadvertent PEEP | < 2 cm H20 |
Inadvertent Continuing Expiratory Pressure | < 2 cm H20 |
Dead Space | < 5.5% of minimum tidal volume |
Peak Inspiratory Flow | 100 lpm for 2 seconds |
Pressure for Initiation of Breath | -2cm H20 maximum |
Operating Conditions | 0° to 122°F |
Storage / Shipping Conditions | -40° to 140°F |
Oxygen Concentration | 100% or 65% with mixing |