Anritsu MA24330A 33 GHz Microwave CW USB Power Sensor
Condition: Refurbished / Calibrated
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- Anritsu MA24330A 33 GHz Microwave CW USB Power Sensor (BC# -L/T*)
Anritsu MA24330A 33 GHz Microwave CW USB Power Sensor
The MA24330A series Microwave CW USB Power Sensors employ a single-path diode architecture to provide fast, accurate average power measurements.
Features:
- Broad Frequency Range Ideal for general purpose, aerospace and defense, satellite and wireless communications applications
- Accurate Power Measurements with over 90 dB Dynamic Range
- Best-in-Class Damage Protection (+26 dBm CW, +32 dBm peak < 10 µs): Protects instrumentation investment
- No Zeroing Required (for signals > –50 dBm) and Elimination of 1 mW Reference Calibration: Reduces test time and handling in production while maintaining absolute accuracy
- Advanced Trigger Capabilities: Facilitates time dependent power measurements
- NIST Traceable Calibration: Provides high-accuracy measurements and ensures absolute accuracy
- Calibration Traceable to SI Units via National Metrology Institutes
- Easy to Use with PC or Select Anritsu Handheld Instruments: No benchtop power meter unit needed
- Silicone Protective Covering (removable): Provides additional field durability
- External Trigger Latching: For pulses as narrow as 20 ns
Specifications:
| Frequency | 10 MHz to 33 GHz, K(m) Connector |
| Power Measurement | Dynamic Range: –70 dBm to +20 dBm Damage Levels at RF Port: +26 dBm, ±20 V DC (+32 dBm peak <10 μs pulse and 10 % duty cycle), minimum |
| Response | Signal Channel Rise Time: 8 μs characteristic Sampling Rate: 140 kS/s |
| Trigger | Source: Bus, Continuous, Internal, External Arm Type (for Internal/External): Auto, Single, Multiple, Standby |
| Internal Trigger | Dynamic Range: –35 dBm to +20 dBm Level Accuracy: ±0.5 dB characteristic Slope: Positive or Negative Delay Range: –5 ms to +10 s Delay Resolution 10 μs Hysteresis: 0 dB to 10 dB, with 0.1 dB resolution Trigger Hold Off: 0 s to 10 s, with 0.01 ms resolution |
| External Trigger | External Trigger Input: MCX (female), 5.5 V maximum Input Impedance: 4 kΩ nominal Type: TTL/CMOS Slope: Positive or Negative Delay Range: –5 ms to +10 s Delay Resolution: 10 μs High Level Input Voltage: 2.3 V min, 3.0 V max Low Level Input Voltage: 1.3 V min, 1.6 V max Latency: 7.1 μs max Trigger Pulse Width: 20 ns min Trigger Repetition Period: 7.1 μs min Trigger Holdoff: 0 s to 10 s with 0.01 ms resolution |
Sell Your Equipment
At AAATesters we are committed to helping you maximize the value of your surplus, excess and underutilized Test Equipment. We buy Test Equipment of all makes and models.
If you would like to Sell your Equipment, please submit the Request form below.
Reasons to Sell to Us
• Quick & Easy Process to Sell or Trade Equipment
• Get Cash Today! or Trade-in Credit for New or Refurbished Equipment
• Free Shipping - We will Pay for Shipping to us. We can even setup, pack and pickup the Equipment you are selling to us
• Improve ROI, Operational Efficiency, Budget Management, and Planning of your Future Needs