Fluke 718 30G
The complete solution to pressure calibration
The Fluke 718 pressure calibrator family (includes 4 models) provide a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.
Features:
- Pressure Source and Milliamp Measurement to calibrate and maintain almost any pressure device
- Integrated pump is easily cleaned when accidently exposed to fluids that reduces cost of ownership and repairs and enables servicing the pump in the field
- Error % calculation for quicker pass/fail decision making in the field
- Best in class mA accuracy of 0.015 % for better measurement confidence and workload coverage
- Min/Max/Hold to capture changing measurements
- Switch test feature for quicker and more accurate pressure switch testing
- New 1 and 300 PSI pressure ranges mean fewer extra tools required
- 1, 30, 100 & 300 PSI ranges now available
Specifications:
Technical Specifications |
Standard |
718 Series |
|
Measure Pressure |
(internal sensor) |
Accuracy |
0.05% full scale |
Functions |
zero, min/max, hold, damp |
Note: |
Dry air, non-corrosives only |
|
|
718 30G |
|
Range: |
-12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa) |
Resolution: |
0.001 psi, 0.1 mbar, 0.01 kPa |
Over Pressure: |
5x Full Scale |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Fluke 718 series via using 700 Series Pressure Modules |
Range: |
29 pressure modules, 0 ヨ 1 in H20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs |
Resolution: |
Per pressure module specs |
Accuracy: |
To 0.025 % of full span, per pressure module specs |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: media compatibility per pressure module specs |
|
Fluke 718 series via built in pump |
Range: |
-12 PSI or -850 mbar to full scale Supported Pressure Units; psi, in H20 (4 °C), in H20 (20 °C), cm H20 (4 °C), cm H20 (20 °C), bar, mbar, kPa, inHg, mmHg, kg/cm |
Loop Power: |
Range: 24 V dc |
|
Accuracy: +/- 10 % |
|
Note: Drive; 20 mA into 1000 O for battery > 6.8V; 700 O for battery 5.8 to 6.8 V |
mA Measure: |
Range: 0 mA to 24 mA |
|
Resolution: 0.001 mA |
|
Accuracy: 0.015 % + 1 count |
|
General Specifications |
Maximum voltage |
|
Temperature |
|
Operating Temperature |
|
Relative Humidity |
|
95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C) |
|
Operating Altitude |
|
Shock |
|
Vibration |
|
Random, 2 g, 5 Hz ヨ 500 Hz |
|
Safety |
|
CSA C22.2 No. 1010.1: 1992 |
|
EMC |
|
EN50082-1: 1992 and EN55022: 1994 Class B |
|
Size |
|
216 mm x 94 mm x 66 mm (8.50 in x 3.72 in x 2.60 in) |
|
Weight |
|
Power, battery life |
|
9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718 |
|
Battery Life |
|
4 to 20 hours typical, depending on functions used |
|
Display |
|
LCD, 5 digit pressure and current simultaneous |
|
Note |
|
For use with non-corrosive gasses |
|
Fluke 700P01
Features:
- Display pressure readings in any of 10 different pressure units you specify in the calibrator setup
- Rugged urethane molded case protects the module from rough handling and harsh conditions
- Features internal temperature compensation from 0ºC to 50ºC for full-accuracy performance
- Includes NIST-traceable calibration certificate
- Module can be calibrated locally using the Fluke 700PCK Pressure Calibration Kit, a PC, and a precision pressure source
The Fluke 700 Series Pressure Modules allow you to measure pressure with the Fluke 700 Series Process Calibrators. Read this sheet before you use the pressure module. This sheet contains specifications and information about how to avoid damaging the pressure module. See the Calibrator Users Manual for operating instructions.
The pressure module measures pressure using an internal microprocessor. It receives operating power from and sends digital information to the 700 Series calibrator. Gage pressure modules have one pressure fitting and measure pressure with respect to atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting. A differential pressure module functions like a gage module when the low fitting is open.
Specifications:
- Range: 10.00 in H2O / 2.5 kPa
- Resolution: 0.01 / 0.002
- Reference Uncertainty: 0.200%
- Total Uncertainty 1 year: 0.300
- High Side Media: Dry
- Low Side Media: Dry
Fluke 700ILF
The Fluke 700ILF In-line Filter is specifically designed to trap fine particles from atmosph ric air, compressed air, and instrument air. It is used with the Fluke 713, 717, or 718 Pressure Calibrators. It can also be used to absorb small quantities of liquid oil and water in droplet form. The filter has the following specifications:
- Inlet fitting: 1/8 in. brass female NPT
- Outlet fitting: 1/8 in. brass male NPT
- Length: 2.125 in.
- Flange diameter: 1.875 in.
- Housing diameter: 1.560 in.
- Max pressure: 100 psi
- Max pressure: 100 psi
- Max flow: 10 SCFM
- Burst pressure: 375 psi
- Filtration: 1 micron
Includes:
- Fluke 718 30G Pressure Calibrator
- Fluke 700P01 Pressure Module w/ Instuction Sheet
- Fluke 700ILF In-Line Filter w/ Instuction Sheet
- Fluke Test Lead Set (1 Red - 1 Black)
- Fluke Test Pins (1 Red - 1 Black)
- Fluke 700P001 Extension Fitting
- Carrying Strap
- User Manual On CD-ROM
- Hard Copy User Manual
- Soft Carrying Case
- Original Factory Box For 700P01 Module
For full Fluke 718 specifications, please click here: Fluke 718
Fluke 718 30G Pressure Calibrator With 700P01 Pressure Module & 700ILF In-Line Filter
The Fluke 718 30G Unit Is Like New & Has Never Been Used!!
700P01 Pressure Module:
Range: 0 to 10.00 in H2O / 2.5 kPa
Resolution: 0.01 / 0.002
Includes Carrying Case & Accessories As Listed!
Fluke 718 30G 71830G Pressure Calibrator Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.