Fluke Hart 1524 Dual Channel Reference Thermometer
Condition: Refurbished / Calibrated
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- Fluke 1524 Reference Thermometer (BC# 20718-L/T*)
- DC Car Charger
- AC Adapter/Charger
- Power Cord
- User Manual On CD-ROM
- Carrying Case (G)
Fluke 1524 Reference Thermometer
Finally, a reference thermometer as versatile as you are
The new 1524 Reference Thermometer from Flukeメs Hart Scientific division measures, graphs, and records PRTs, thermocouples, and thermistors. The new thermometer readouts deliver high accuracy, wide measurement range, logging, and trending, all in a handheld tool you can take anywhere.
The 1524 lets you handle field applications, laboratory measurements, and data logging with ease. And with the dual channel measurement capabilities of the model 1524, you can do twice the work in half the time.
Make accurate, consistent measurement...anywhere
You need accuracy for compliance, product yields, energy savings, and consistent results. The 1524 uses current reversal, a technique used in high-end instruments that eliminates thermal EMFs, for precision temperature measurements. Specifications are guaranteed from ヨ10 °C to 60 °C ambient. Special precision resistors and a highly stable reference voltage source keep 1524 accuracy virtually insensitive to environmental temperature.
Like all Fluke handheld tools, the 1524 Reference Thermometer endures rigorous testing in temperature extremes and under harsh conditions of vibration, so you can take it with confidence anywhere you need to go.
Monitor trends in the lab or in the field
See trends graphically on the 1524 thermometerメs 128x64-backlit LCD display. You can change the graphメs resolution at the touch of a button. Now it is easy to see when the temperature is stable (without statistics or long delays) or to monitor processes over time to verify correct operation.
Hold readings on the display at the touch of a button, or document up to 25 readings and associated statistics for easy retrieval. Statistics include the average maximum and minimum values, and the standard deviation.
View them through the display or by uploading it to a PC via RS-232 conection and 9940 I/O TookKit software, included free. To monitor and log more data over time, use a PC and optional LogWare II software
RS-232-to-USB adapters are available for those who prefer USB connectors. Battery power lasts at least 20 hours on three AA batteries, or use the dc power adapter for extended periods of measurement. Power saving features can be enabled or disabled for longer battery life or greater convenience.
INFO-CON connectors ensure correct temperature conversion
Inside the INFO-CON, a memory chip keeps calibration information for the attached probe. Simply plugging in the probe uploads the information to the readout. The connector transfers this information to the 1524 automatically, ensuring the correct temperature conversion for accurate, hassle-free measurements
Probes may be locked by password to specific channels and readouts for security or for system calibration traceability. Plug any thermocouple with mini-thermocouple jacks into an optional universal thermocouple adapter for convenient measurement. Each thermocouple adapter or standard connector supports reference junction compensation (RJC) with its own internal precision thermistor.
Features:
- High accuracy
- PRTs: up to ± 0.011 °C
- Thermocouples: up to ± 0.24 °C
- Precision thermistors: ± 0.002 °C
Specifications:
Input channels | 2 |
Resolution | Resolution PRTs and thermistors: 0.001° thermocouples: 0.01° |
Logging | 25 readings with statistics15,000 time and date stamped |
Sample interval (normal) | 1 second (simultaneous measurement) |
Typical sample interval (fast mode)* | 0.3 seconds |
Sensor types | PRTs, RTDs, thermistors, and thermocouples |
Thermocouple types | C, E, J, K, L, M, N, T, U, B, R, S |
Statistics | Maximum, minimum, average, standard deviation |
Trending | Scale: ± 10 °C (18 °F), ± 1 °C (1.8 °F), ± 0.1°C (0.18 °F), ± 0.01°C (0.018 °F), 10 minutes of real-time data |
Power requirements | 3 AA alkaline batteries, 12 V dc universal power supply |
Size (HxWxD) | 96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 in) |
Weight | 0.65 kg (1.4 lb) |
Computer interface | RS-232, 9940 I/O ToolKit software included |
Safety | EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |
Environmental conditions for best accuracy: 13 °C to 33 °C (55.4 °F to 91.4 °F) | |
Millivolt range and accuracy | ヨ10 mV to 75 mV, ± (0.005 % + 5 µV) |
Internal reference junction compensation | ± 0.2 °C (± 0.36 °F) |
Resistance range and accuracy | 0 O to 400 O ± (0.004 % + 0.002 O) 200 O to 50 kO ± (0.01 % + 0.5 O) 50 kO to 500 kO ± (0.03 %) |
Temperature coefficient, voltage: ヨ10 °C to 13 °C , +33 °C to 60 °C (14 °F to 55.4 °F, 91.4 °F to 140 °F) |
± (0.001 %/°C + 1 mV/°C) |
Temperature coefficient, resistance: ヨ10 °C to 13 °C , +33 °C to 60 °C (14 °F to 55.4 °F, 91.4 °F to 140 °F) |
0.0008 %/°C + 0.0004 O (0 O to 400 O) 0.002 %/°C + 0.1 O (0 O to 50 kO) 0.06 %/°C + 0.1 O (50 kO to 500 kO) |
Excitation current, resistance | 1 mA (0 O to 400 O) 10 µA (0 O to 50 kO) 2 µA (50 kO to 500 kO) |
Equivalent temperature accuracies for selected sensors derived from primary specifications (O, mV)
Temperature, Thermocouples:
Type | Range | Measure accuracies |
K | ヨ200 °C to 0 °C (-328 °F to 32 °F) | ± 0.61 °C (± 1.10 °F) |
0 °C to 1370 °C (32 °F to 2498 °F) | ± 0.24 °C (± 0.43 °F) | |
R | ヨ20 °C to 0 °C (4 °F to 32 °F) | ± 1.09 °C (± 1.96 °F) |
0 °C to 500 °C (32 °F to 932 °F) | ± 0.97 °C (± 1.71 °F) | |
500 °C to 1750 °C (932 °F to 3182 °F) | ± 0.49 °C (± 0.88 °F) | |
S | ヨ20 °C to 0 °C (4 °F to 32 °F) | ± 1.05 °C (± 1.89 °F) |
0 °C to 500 °C (32 °F to 932 °F) | ± 0.95 °C (± 1.71°F) | |
500 °C to 1750 °C (932 °F to 3182 °F) | ± 0.56 °C (± 1.01 °F) | |
Accuracies are based on internal reference junction compensation. Refer to technical manual for greatly improved accuracies using external reference junctions. |
Accuracies of selected readout/probe combinations (±°C)
Temperature | 5616-12 | 5615-6 | 5627A-12 | 5610-9 |
-200 °C (-328 °F) | 0.014 | 0.025 | 0.027 | n/a |
0 °C (32 °F) | 0.021 | 0.021 | 0.049 | 0.009 |
100 °C (212 °F) | 0.027 | 0.028 | 0.065 | 0.009 |
300 °C (572 °F) | 0.040 | 0.043 | 0.103 | n/a |
420 °C (788 °F) | 0.050 | n/a | 0.130 | n/a |
Includes readout accuracy, probe calibration, and probe drift |
For full Fluke 1524 specifications, please click here: Fluke 1524