Features & Applications:
Modes
The Fluke 96B features three modes of operation. The first is multimeter mode. This displays standard multimeter information, including voltage, amperage and resistance. Multimeter mode is often used for troubleshooting simple electronic circuits. The second available option is oscilloscope mode. In this mode, the Fluke 96B displays a digital waveform. The digital oscilloscope mode is useful for analyzing complex digital circuits. A third mode combines the previous two modes and displays scope and multimeter functions on one screen.
Power
The Fluke 96B comes with a standard internal battery pack, which allows for portable operation. The battery pack contains nickel-cadmium (NiCAD) 4.8 volt power cells and can power the Fluke for up to four hours. The Fluke 96B can also be powered from an external source, using an 8 volt to 20 volt direct current (DC) power supply with a 5 mm plug.
Safety
The Fluke 96B is designed to meet International Electrotechnical Commission (IEC) safety standards and is UL-listed for electrical measurements up to 600 volts. The device features overload protection that stops electrical flow if 300 volts or more are detected. The Fluke also has surge protection that can resist electrical spikes up to four kilovolts.
Display
The display of the Fluke 96B is a digital monochrome screen with a zoom function. This zoom allows users to quickly focus on specific details of the oscilloscope display. The Fluke 96B oscilloscope screen play is capable of displaying data at 2.5 giga-samples per second (GS/s).
SCOPE MODE
- Continuous Auto Set
Hands-free signal probing. Continuously follows input
signals from 15 Hz to 50 MHz and automatically selects the
proper time base, input range, and trigger level. - Min Max Envelope
Records the waveform excursions with a grey scale
envelope on both inputs. - 40 ns Glitch detect
Sweep speeds 1 µs per division or slower on INPUT A. - Grey-scale display
Analog-like Grey-scale display. - Waveform Average
Smoothing up to 256 waveforms. 10 samples in Roll mode.
PowerAdvanced METER Functions
Touch Hold®, Change Alertル, Smoothingル, Continuity Beeper. Readouts in % change, % scale or delta-zero.
MIN MAX Trend Plot
Chart recording function to log MIN, MAX, and AVERAGE readings. Displays all 3 readings as graphs with time and date stamp. Automatic vertical scaling and horizontal time compression for recording from 120 seconds to 40 days.
Vertical | |
Bandwidth | 60 MHz |
Coupling | AC, DC, Ground |
Sensitivity | 1 mV to 100 V/div (With included 10:1 probe measure up to 600 Vrms) |
Modes | A, -A, B, -B, A+B, A-B, X vs. Y |
Vertical resolution | 8 bits (256 levels) |
Accuracy | ¡¾(2% of reading +1 pixel) |
Horizontal | |
Range | 5 ns/div to 60 s/div (Fluke 105B and Fluke 99B) 10 ns/div to 60 s/div (Fluke 96B and Fluke 92B) |
Acquisition modes | Recurrent, Single shot, Envelope, Roll, Glitch detect, Average, ScopeRecordル |
Maximum sample rate | 2.5 GS/s for repetitive signals; 25 MS/s for single shot |
Maximum record length | Scope Recordル 30k (up to 20ms/div-1 ch) 15k (up to 20ms/div-2 ch) 512 samples per channel (up to 5ns/div) |
Accuracy | ¡¾(-0.1% of reading +1 pixel) |
Trigger | |
Source | Input A, Input B, External |
Modes | Free run, Triggered, Single |
Types | Edge, Video, Event, N-cycle |
Video | Triggers on Field 1, Field 2, Lines or selectable line number |
Interlaced video | NTSC, PAL, PAL/M, SECAM, User selectable 2-Field or 4-Field sequence |
Non Interlaced video | Line selection up to 3200 lines |
Line frequency | 14 kHz to 65 kHz |
Time delay | Up to 20 divisions pre trigger view Up to 640 divisions trigger delay |
Cursor Measurements | Up to 5 displayed simultaneously, Vrms (AC), Frequency, Vmean (DC), Risetime, Vpeak-peak, Time between cursors, Vmax-peak, Time at cursors (Time of day or Time from start), Vmin-peak, Phase dV between cursors, V at cursors. Measurements can be scaled to Amperes. Readouts in % change or delta-zero. |
Zoom | Adjustable magnification. Allows full 30k record overview, or zoom in to single sample resolution. |
Memory Save and Recall | |
Setup Memories | 20 |
Waveform Memories | 10 |
Screen Memories | 5 |
DC Voltage | |
Ranges | 100 mV to 300V (With included 10:1 probe measure up to 600 Vrms) |
Accuracy | ¡¾(0.5%+5 counts) |
AC or AC+DC true rms Voltage | |
Ranges | 100 mV to 250V (with included 10:1 probe measure up to 600 Vrms) |
Accuracy | 50 Hz to 60 Hz ¡¾(1%+10 counts) 20 Hz to 20 kHz ¡¾(2%+15 counts) 5 Hz to 1 MHz ¡¾(3%+20 counts) DC to 5 MHz ¡¾(10%+25 counts) |
Additional Measurements | Continuity, Diode test, Amps, Hz, ¡ÆC or ¡ÆF, Pulse Width, Duty Cycle, RPM, dBm, dBV, dBW, Audio or RF Watts. (Amps, ¡ÆC or ¡ÆF with Optional Probes) |
General Specifications | |
Display | Cold cathode fluorescent back-lit liquid crystal |
Size | 84x84 mm (3.3x3.3 inch) |
Contrast and brightness | User adjustable |
Real-time clock | Stamps Min Max TrendPlots and 30k records with time and date. |
Power | Line voltage adapter/battery charger included Installed battery Rechargeable NiCad Pack Operating time 4 hours typical Charging time 21 hours typical Alternate battery 4 Alkaline C cells (user supplied) (non rechargable)Environmental |
Mechanical Data | |
Size | 60x130x260 mm (2.4x5.1x10.2 inches) |
Weight | 1.5 kg (3.3lbs) |