Fluke Biomedical ProSim 4 & SPOT Light Vital Signs Simulator SpO2 Tester

Fluke Biomedical ProSim 4 & SPOT Light Vital Signs Simulator SpO2 Tester

Fluke Biomedical ProSim 8 Vital Signs & ECG Patient Simulator with ProSim Spot Sp02 Test Module

Fluke Biomedical ProSim 8 Vital Signs & ECG Patient Simulator with ProSim Spot Sp02 Test Module

Fluke Biomedical ProSim 4 Vital Signs & ECG Patient Simulator

$4,495.00
Availability: Out of stock
SKU
MT56

Condition: Refurbished / Calibrated

Warranty: 1 Year Warranty

Request a Quote for

Have Questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

we are commited to win your business!

Fluke Biomedical ProSim 4 Vital Signs Simulator

ProSim 4 Vital Signs Simulator with breakthrough touchscreen technology offers quick and simple one-tap testing for patient monitor performance checks and troubleshooting. Designed to get you in and out of most locations in 60 seconds, this quick-check device offers 12-lead ECG simulation, respiration, IBP and NIBP testing in the palm of your hand. Featuring specialized stay-connected ECG posts to ensure secure lead connections and no-hassle testing, ProSim 4 is the perfect patient simulator for first-call patient monitor quality assurance and safety professionals.

 

Features:

  • Portable multifunction tester offers 12-lead ECG, respiration, IBP and NIBP simulation
  • 90 % smaller and lighter than combined technology of legacy products
  • Breakthrough touchscreen technology
  • One-tap testing for most performance tests and checks
  • Easy quick-check patient monitor testing in one minute or less with onboard, customizable patient pre-sets and autosequences
  • Integrated, easily-replaceable battery capable of running quick checks all day
  • Stay-connected ECG posts for secure lead connections
  • Repeatable NIBP testing within 2 mmHg independent of device under test
  • Multi-language user interface offers choice of language selection
  • Tilt stand design for operation in tight spaces and better viewing angle

 

Specifications:

General
Temperature Operating 10 °C to 40 °C (50 °F to 104 °F)
Storage -20 °C to +60 °C (-4 °F to + 140 °F)
Humidity 10 % to 90 % non-condensing
Altitude 3,000 meters (9,843 ft)
Dimensions 18 cm x 9.3 cm x 5.5 cm (7.1 in x 3.7 in x 2.2 in)
Display LCD touch-screen color display
Communication USB port (for calibration and firmware updates only)
Power Lithium-ion rechargeable battery
Battery charger 110 to 220 V, 50/60 Hz input, 6 V/3.5 A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle
Weight Four hours (minimum), 40 NIBP cycles typical
Battery life 0.88 kg (1.93 lb)
Safety standards IEC 61010-1:2001
Certifications CE, CSA, C-TICK N10140, RoHs
EMC IEC 61326-1:2006
Detailed specifications
Normal-sinus-rhythm waveform
ECG reference The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional
Normal sinus rhythm 12-lead configuration with independent outputs referenced to right leg (RL). Output to 10 universal ECG Jacks, color-coded to AHA and IEC standards
Amplitude 1.0 mV. Other leads are proportional to Lead II (reference lead) in percentage per: Lead I: 70 Lead II: 100 Lead III: 30 Lead V1: 24 Lead V2: 48 Lead V3: 100 Lead V4: 120 Lead V5: 112 Lead V6: 80
Amplitude accuracy ± 5 % of setting Lead II
ECG rate 30 BPM, 60 BPM, 80 BPM, 90 BPM, 120 BPM, 150 BPM, 180 BPM, 210 BPM, 240 BPM, 270 BPM, 300 BPM, and 320 BPM. Preset and monitor testing sequence hypotensive condition is at 40 BPM
Rate accuracy ± 1 % of setting
ECG waveform slection Adult (80 ms) or neonatal (40 ms) QRS duration
Power-on default 60 BPM, 1.0 mV, adult QRS
Arrhythmia
Atrial fibrillation Coarse or fine
Premature ventricular contraction Left ventricular
Ventricular tachycardia 160 BPM or 200 BPM
Ventricular fibrillation Coarse or fine
Transvenous pacer pulse 75 BPM, left arterial, 3 mV amplitude on lead II, accuracy ± 10 %, 1.0 ms width
2nd degree AV block Type 1
3rd degree AV block 3rd degree AV block
Asystole Asystole
ECG performance testing
Amplitude 1 mV. Other leads are proportional to Lead II (reference lead) in percentage per: Lead I: 70 Lead II: 100 Lead III: 30 Lead V1: 24 Lead V2: 48 Lead V3: 100 Lead V4: 120 Lead V5: 112 Lead V6: 80
Square wave 60 ms at 2 Hz
Respiration
Rate 0 (OFF), 10 BrPM to 100 BrPM in 10 BrPM steps
Impedance variations 1 Ω
Accuracy delta ± (10 % + 0.05 ohm)
Baseline 500 Ω to circuit common, giving 1000 Ω between any two leads
Accuracy baseline ± 5 %
Respiration lead LA or LL (default)
Invasive blood pressure
Channels 1 electrically isolated from all other signals
BP output Circular DIN 5-pin
Input/output impedance 300 Ω ± 10 %
Exciter input range 2 to 16 V peak
Exciter-input frequency range DC to 5000 Hz
Transducer sensitivity 5 µV/V/mmHg
Pressure accuracy ± (1% of setting + 1 mmHg) Accuracy guaranteed for dc excitation only
Static pressure 0 mmHg, 80 mmHg, 160 mmHg, and 250 mmHg
Dynamic waveforms Synchronization To ECG heartrate Chambers simulated and systolic/diastolic pressure:
Type IBP (arterial) IBP (left ventrical)
Adult IBP (arterial) 60/30 IBP (left ventrical) 60/0
Adult IBP (arterial) 120/80 IBP (left ventrical) 120/0
Adult IBP (arterial) 150/100 IBP (left ventrical) 150/0
Adult IBP (arterial) 200/150 IBP (left ventrical) 200/0
Neonatal IBP (arterial) 35/15 IBP (left ventrical) 35/0
Neonatal IBP (arterial) 70/40 IBP (left ventrical) 70/0
Non-invasive blood pressure
Pressure units mmHg
Manometer Range: 10 mmHg to 400 mmHg
Resolution: 0.1 mmHg (for display purposes)
Accuracy: ± (1 % reading + 1 mmHg)
Pressure source Inflation bulb or device under test
NIBP simulations Pulse: 2 mmHg max into 500 ml NIBP system
Volume of air moved: 1 ml max
Simulations: Adult: 60/30 (40), 120/80 (93); 150/100 (117); and 200/150 (167) Neonatal: 35/15 (22) and 70/40 (50)
Repeatability: Within ± 2 mmHg (at maximal pulse size independent of device under test)
Synchonization: To ECG heartrate (maximal rate 120 BPM)
Leak test Target pressure: 20 mmHg to 400 mmHg
Elapse time: 0:30 minutes to 5:00 minutes: seconds in 30 second steps
Leakage rate: 0 to 200 mmHg/minute
Pressure relief test range 100 mmHg to 400 mmHg

Fluke Biomedical ProSim 4 Vital Signs Simulator

Designed to get you in and out of most locations in 60 seconds, this quick-check device offers 12-lead ECG simulation, respiration, IBP and NIBP testing in the palm of your hand

Includes Carrying Case & Accessories As Listed

Pro sim 4 

Fluke Biomedical ProSim 4 ProSim4 Vital Signs Simulator Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.

 

BC# 7514A-L/T*

  • Fluke Biomedical ProSim 4 Vital Signs Simulator (BC# 7514A-L/T*)
  • AC Adapter/Charger
  • User Manual On CD-ROM
  • Hard Carrying Case