Fluke Biomedical ProSim 4 Vital Signs & ECG Patient Simulator
    
        $5,999.00
                
        
    
            Availability:
            In stock
        
    
            SKU
        
	
	MT56
Condition: Refurbished / Calibrated
Warranty: 1 Year Warranty
Shipping: USA & Worldwide
- Fluke Biomedical ProSim 4 Vital Signs Simulator (BC# 7514A-L/T*)
- AC Adapter/Charger
- User Manual On CD-ROM
- Hard Carrying Case
			Request a Quote for 
	
	
	
Fluke Biomedical ProSim 4 Vital Signs Simulator
ProSim 4 Vital Signs Simulator with breakthrough touchscreen technology offers quick and simple one-tap testing for patient monitor performance checks and troubleshooting. Designed to get you in and out of most locations in 60 seconds, this quick-check device offers 12-lead ECG simulation, respiration, IBP and NIBP testing in the palm of your hand. Featuring specialized stay-connected ECG posts to ensure secure lead connections and no-hassle testing, ProSim 4 is the perfect patient simulator for first-call patient monitor quality assurance and safety professionals.
Features:
- Portable multifunction tester offers 12-lead ECG, respiration, IBP and NIBP simulation
- 90 % smaller and lighter than combined technology of legacy products
- Breakthrough touchscreen technology
- One-tap testing for most performance tests and checks
- Easy quick-check patient monitor testing in one minute or less with onboard, customizable patient pre-sets and autosequences
- Integrated, easily-replaceable battery capable of running quick checks all day
- Stay-connected ECG posts for secure lead connections
- Repeatable NIBP testing within 2 mmHg independent of device under test
- Multi-language user interface offers choice of language selection
- Tilt stand design for operation in tight spaces and better viewing angle
Specifications:
| General | |
| Temperature | Operating 10 °C to 40 °C (50 °F to 104 °F) Storage -20 °C to +60 °C (-4 °F to + 140 °F) | 
| Humidity | 10 % to 90 % non-condensing | 
| Altitude | 3,000 meters (9,843 ft) | 
| Dimensions | 18 cm x 9.3 cm x 5.5 cm (7.1 in x 3.7 in x 2.2 in) | 
| Display | LCD touch-screen color display | 
| Communication | USB port (for calibration and firmware updates only) | 
| Power | Lithium-ion rechargeable battery | 
| Battery charger | 110 to 220 V, 50/60 Hz input, 6 V/3.5 A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle | 
| Weight | Four hours (minimum), 40 NIBP cycles typical | 
| Battery life | 0.88 kg (1.93 lb) | 
| Safety standards | IEC 61010-1:2001 | 
| Certifications | CE, CSA, C-TICK N10140, RoHs | 
| EMC | IEC 61326-1:2006 | 
| Detailed specifications | |
| Normal-sinus-rhythm waveform | |
| ECG reference | The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional | 
| Normal sinus rhythm | 12-lead configuration with independent outputs referenced to right leg (RL). Output to 10 universal ECG Jacks, color-coded to AHA and IEC standards | 
| Amplitude | 1.0 mV. Other leads are proportional to Lead II (reference lead) in percentage per: Lead I: 70 Lead II: 100 Lead III: 30 Lead V1: 24 Lead V2: 48 Lead V3: 100 Lead V4: 120 Lead V5: 112 Lead V6: 80 | 
| Amplitude accuracy | ± 5 % of setting Lead II | 
| ECG rate | 30 BPM, 60 BPM, 80 BPM, 90 BPM, 120 BPM, 150 BPM, 180 BPM, 210 BPM, 240 BPM, 270 BPM, 300 BPM, and 320 BPM. Preset and monitor testing sequence hypotensive condition is at 40 BPM | 
| Rate accuracy | ± 1 % of setting | 
| ECG waveform slection | Adult (80 ms) or neonatal (40 ms) QRS duration | 
| Power-on default | 60 BPM, 1.0 mV, adult QRS | 
| Arrhythmia | |
| Atrial fibrillation | Coarse or fine | 
| Premature ventricular contraction | Left ventricular | 
| Ventricular tachycardia | 160 BPM or 200 BPM | 
| Ventricular fibrillation | Coarse or fine | 
| Transvenous pacer pulse | 75 BPM, left arterial, 3 mV amplitude on lead II, accuracy ± 10 %, 1.0 ms width | 
| 2nd degree AV block | Type 1 | 
| 3rd degree AV block | 3rd degree AV block | 
| Asystole | Asystole | 
| ECG performance testing | |
| Amplitude | 1 mV. Other leads are proportional to Lead II (reference lead) in percentage per: Lead I: 70 Lead II: 100 Lead III: 30 Lead V1: 24 Lead V2: 48 Lead V3: 100 Lead V4: 120 Lead V5: 112 Lead V6: 80 | 
| Square wave | 60 ms at 2 Hz | 
| Respiration | |
| Rate | 0 (OFF), 10 BrPM to 100 BrPM in 10 BrPM steps | 
| Impedance variations | 1 Ω | 
| Accuracy delta | ± (10 % + 0.05 ohm) | 
| Baseline | 500 Ω to circuit common, giving 1000 Ω between any two leads | 
| Accuracy baseline | ± 5 % | 
| Respiration lead | LA or LL (default) | 
| Invasive blood pressure | |
| Channels | 1 electrically isolated from all other signals | 
| BP output | Circular DIN 5-pin | 
| Input/output impedance | 300 Ω ± 10 % | 
| Exciter input range | 2 to 16 V peak | 
| Exciter-input frequency range | DC to 5000 Hz | 
| Transducer sensitivity | 5 µV/V/mmHg | 
| Pressure accuracy | ± (1% of setting + 1 mmHg) Accuracy guaranteed for dc excitation only | 
| Static pressure | 0 mmHg, 80 mmHg, 160 mmHg, and 250 mmHg | 
| Dynamic waveforms | Synchronization To ECG heartrate Chambers simulated and systolic/diastolic pressure: | 
| Type | IBP (arterial) IBP (left ventrical) | 
| Adult | IBP (arterial) 60/30 IBP (left ventrical) 60/0 | 
| Adult | IBP (arterial) 120/80 IBP (left ventrical) 120/0 | 
| Adult | IBP (arterial) 150/100 IBP (left ventrical) 150/0 | 
| Adult | IBP (arterial) 200/150 IBP (left ventrical) 200/0 | 
| Neonatal | IBP (arterial) 35/15 IBP (left ventrical) 35/0 | 
| Neonatal | IBP (arterial) 70/40 IBP (left ventrical) 70/0 | 
| Non-invasive blood pressure | |
| Pressure units | mmHg | 
| Manometer | Range: 10 mmHg to 400 mmHg Resolution: 0.1 mmHg (for display purposes) Accuracy: ± (1 % reading + 1 mmHg) | 
| Pressure source | Inflation bulb or device under test | 
| NIBP simulations | Pulse: 2 mmHg max into 500 ml NIBP system Volume of air moved: 1 ml max Simulations: Adult: 60/30 (40), 120/80 (93); 150/100 (117); and 200/150 (167) Neonatal: 35/15 (22) and 70/40 (50) Repeatability: Within ± 2 mmHg (at maximal pulse size independent of device under test) Synchonization: To ECG heartrate (maximal rate 120 BPM) | 
| Leak test | Target pressure: 20 mmHg to 400 mmHg Elapse time: 0:30 minutes to 5:00 minutes: seconds in 30 second steps Leakage rate: 0 to 200 mmHg/minute | 
| Pressure relief test range | 100 mmHg to 400 mmHg | 
 
     
     
        