Fluke ProSim 8 Vital Signs Simulator

The 8-in-1 ProSim 8 Vital Signs Simulator offers fast and comprehensive preventative maintenance (PM) testing for your entire patient monitor fleet. Designed to get you in and out of most PM locations in minutes, this multifunction simulator tests ECG (including fetal ECG and arrhythmias), respiration, temperature, IBP, cardiac output, NIBP, SpO2, and is capable of testing Rainbow multi-wavelength waveforms. Featuring specialized stay-connected ECG posts for secure lead connections, physiologically-synchronized pulses across all parameters, and customizable patient pre-sets and autosequences, the ProSim 8 patient simulator provides unbeatably fast and easy complete monitor testing. Barcode-scanner compatibility and wireless PC interface, direct printing, data transfer and reporting, along with advanced, integrated technologies and works-every-time performance allow top confidence in patient monitor fleet performance and supports passing regulatory audits with ease.

 

Key Features:

  • All-in-one complete monitor testing 80 % smaller and 17 lbs/7.7 kilos lighter than predecessor technology
  • 8-in-1 multifunction simulator tests ECG (including fetal ECG and arrhythmias), respiration, temperature, IBP, cardiac output, NIBP, SpO2, and Rainbow multi-wavelength waveforms
  • Stay-connected ECG posts for easy/secure ECG snap and lead connections
  • Custom SpO2 r-curve for accurate testing of the latest and future oximetry technologies
  • Static pressure linearity testing
  • Repeatable NIBP simulation (+/- 2 mmHg) for dynamic pressure repeatability testing
  • Physiologically synchronized pulses across all parameters
  • Barcode scanning and direct data capture and printing functionality
  • Onboard, customizable patient pre-sets and autosequences for fast/easy testing
  • Multi-language user interface offers choice of language selection
  • Integrated, easily-replaceable long-life battery

 

Specifications:

General Specifications
Temperature Operating: 10 °C to 40 °C (50 °F to 104 °F)
Storage: -20 °C to +60 °C (-4 °F to 140 °F)
Humidity 10 % to 90 % non-condensing
Altitude 3,000 meters (9,843 ft)
Dimensions (L x W x H) 14.5 cm x 30.2 cm x 8.6 cm (5.7 in x 11.9 in x 3.4 in)
Display LCD color display
Communication USB device upstream port: Mini-B connector for control by a computer
USB host controller port: Type A, 5 V output, 0.5 A max load. Connector for keyboard, barcode reader, and printer
Wireless: IEEE 82.15.4 for control by a computer
Power Lithium-ion rechargeable battery
Battery charger 100 V to 240 V input, 15 V/2.0 A output. For best performance, the battery charger should be connected to a properly-grounded ac receptacle
Battery life 9 hours (minimum), 100 NIBP cycles typical
Weight 1.87 kg (4.2 lb)
Safety standards EN/IEC 61010-1:2001
Certifications CE, CSA, C-TICK N10140 , RoHS
Electromagnetic compatibility (EMC) IEC 61326-1:2006
Detailed Specifications
Normal-sinus-rhythm waveform
ECG reference The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional
Normal sinus rhythm 12-lead configuration with independent outputs referenced to right leg (RL). Output to 10 universal ECG jacks, color-coded to AHA and IEC standards
High-level output 0.5 V/mV ± 5 % of the ECG amplitude setting available on a BNC connector
Amplitude 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV steps) Other leads are proportional to Lead II (reference lead) in percentage per:
Lead I: 70
Lead II: 100
Lead III: 30
Lead V1: 24
Lead V2: 48
Lead V3: 100
Lead V4: 120
Lead V5: 112
Lead V6: 80
Amplitude accuracy ± (2 % of setting + 0.05 mV)
ECG rate 10 BPM to 360 BPM in 1 BPM steps
Rate accuracy ± 1 % of setting
ECG waveform selection Adult (80 ms) or pediatric (40 ms) QRS duration
ST-segment elevation Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps).
Additional steps: + 0.05 mV and - 0.05 mV
Power-on default 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
Pacemaker waveform
Pacer pulse Amplitude: 0 (off), ± 2, ± 4, ± 6, ± 8, ± 10, ± 12, ± 14, ± 16, ± 18, ± 20, ± 50, ± 100, ± 200, ± 500, and ± 700 mV for lead II (reference lead)
Accuracy: Reference lead II: ± (5 % setting + 0.2 mV) All other leads: ± (10 % setting + 0.4 mV)
Pacer pulse width 0.1 ms, 0.2 ms, 0.5 ms, 1 ms, and 2 ms ± 5 %
Paced arrhythmias Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-ventricular sequential
Noncapture (one time)
Nonfunction
Power-on default Amplitude 5 mV, width 1 ms, atrial waveform
Arrhythmia
Baseline NSR 80 BPM
PVC focus Left focus, standard timing (except where specified)
Supraventricular arrhythmia Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed beat (one time); atrial tachycardia; paroxysmal atrial tachcardia; nodal rhythm; and supraventricular tachycardia
Premature arrhythmia Premature atrial contraction (PAC); premature nodal contraction (PNC); PVC1 left ventricular; PVC1 left ventricular, early; PVC1 left ventricular, R on T; PVC2 right ventricular; PVC2 right ventricular, early; PVC2 right ventricular, R on T; and multifocal PVCs
Ventricular arrhythmia PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy; trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); monoventricular tachycardia (120 to 300 BPM in 5 BPM steps); poly-ventricular tachycardia (5 types); ventricular fibrillation (coarse or fine); and asystole
Conduction defect First-, second-, or third-degree heart block; and right- or left-bundlebranch block
Advanced cardiac life support Shockable pulseless arrest rhythms: Ventricular fibrillation (coarse), ventricular fibrillation (fine), unstable polymorphic ventricular tachycardia
Non-shockable pulseless arrest rhythms: Asystole
Symptomatic bradycardia: Sinus bradycardia (< 60 BPM) 2nd degree AV block, mobitz type I 2nd degree AV block, mobitz type II Complete/3rd degree AV block Right bundle branch block Left bundle branch block
Symptomatic tachycardia: regular narrow-complex tachycardia (QRS < 0.12 seconds): Sinus tachycardia > 150 BPM Supraventricular Tachycardia
Symptomatic tachycardia: regular wide-complex tachycardias (QRS ≥ 0.12 seconds): Sinus tachycardia > 150 BPM Supraventricular tachycardia SVT with aberrancy
Irregular tachycardia: Atrial fibrillation (coarse and fine), atrial flutter, unstable monomorphic ventricular tachycardia (120 BPM to 300 BPM), torsade de pointes/polymorphic ventricular tachycardia (long QT interval)
ECG Performance testing
Amplitude 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV steps) Other leads are proportional to Lead II (reference lead) in percentage per:
Lead I: 70
Lead II: 100
Lead III: 30
Lead V1 through V6: 100
Pulse wave 30 BPM, 60 BPM, with 60 ms pulse width
Square wave 0.125 Hz, 2 Hz, 2.5 Hz
Triangle wave 0.125 Hz, 2 Hz, 2.5 Hz
Sine wave 0.05 Hz, 0.5 Hz, 1, 2 Hz, 5 Hz, 10 Hz, 25 Hz, 30 Hz, 40 Hz, 50 Hz, 60 Hz, 100 Hz, and 150 Hz
R-wave detection Waveform: Triangular pulse
Rate: 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
Width: 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
Width accuracy: ± (1 % of setting + 1 ms)
QRS detection Widths 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
Width accuracy ± (1 % of setting + 1 ms)
Rate 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
R-Wave up slope 0.875 amplitude, 0.4375 x width
R-Wave down slope Full amplitude, 0.5 x width
S-Wave up slope 0.125 amplitude, 0.0625 x width
Tall T-wave rejection Waveform QT Interval 350 ms T-Wave width 180 ms T-Wave shape ½ sinewave
Amplitude 0 % to 150 % reference lead amplitude in 10 % steps
Rate 80 BPM
Rate accuracy ± 1 % of setting
Amplitute accuracy ± (2 % of setting + 0.05 mV)
ECG artifact
Type 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead
Lead select All, RA, LL, LA, V1, V2, V3, V4, V5, V6
Fetal/Maternal ECG
Fetal heart rate (fixed) 60 BPM to 240 BPM in 1 BPM steps
Fetal heart rate (IUP) 140 BPM at beginning, then varies with pressure
Intrauterine-pressure waveforms Early deceleration, late deceleration, and acceleration
Wave duration 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0
IUP period 2 min, 3 min, or 5 minutes; and manual
Default settings FHR 140 BPM, early deceleration wave, manual
Invasive blood pressure
Channels 2, each independently settable with identical parameters and are individually electrically isolated from all other signals
Input/output impedance 300 Ω ± 10 %
Exciter input range 2 to 16 V peak
Exciter-input frequency range DC to 5000 Hz
Transducer sensitivity 5 (default) or 40 µV/V/mmHg
Pressure accuracy ± (1 % of setting + 1 mmHg) accuracy guaranteed for dc excitation only
Static pressure - 10 to + 300 mmHg in 1 mmHg steps
Pressure units mmHg or Kpa
Dynamic waveforms Types (default pressures):
Arterial (120/80) Radial artery (120/80) Left ventricle (120/00) Right ventricle (25/00) Pulmonary artery (25/10) Pulmonary-artery wedge (10/2) Right atrium (central venous or CVP) (15/10)
Pressure variability: Systolic and diastolic pressures are independently variable in 1 mmHg steps
Swan-Ganz sequence Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary artery wedge (PAW)
Cardiac catheterization Chambers: Aortic, pulmonary valve, and mitral valve
Respiration artifact Arterial, radial artery, and left ventricle: 5 % to 10 % multiplication
Other: 5 mmHg or 10 mmHg
BP output Circular DIN 5-Pin
Power-on default 0 mmHg
Respiration
Rate 0 (OFF), 10 BrPM to 150 BrPM in 1 BrPM steps
Waves Normal or ventilated
Ratio (inspiration:expiration) Normal 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated 1:1
Impedance variations (Δ Ω) 0.00 Ω to 1.00 Ω in 0.05 Ω steps and 1 Ω to 5 Ω in 0.25 Ω steps
Accuracy delta ± (5 % of setting + 0.1 Ω)
Baseline 500 Ω, 1000 Ω (default), 1500 Ω, 2000 Ω, Leads I, II, III
Accuracy baseline ± 5 %
Respiration lead LA or LL (default)
Apnea selection 12 sec, 22 sec, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
Power-on default 20 BrPM, delta 1.0 Ω
Temperature
Temperature 3.0 °C to 42.0 °C in 0.5 °C steps
Accuracy ± 0.4 °C
Compatibility Yellow Springs, Inc. (YSI) Series 400 and 700
Output Circular DIN 4-Pin
Cardiac output
Catheter type Baxter Edwards, 93a-131-7f
Calibration coeffecient 0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood temperature 36 °C (98.6 °F) to 38 °C (100.4 °F) ± 0.2 °C in 1 °C steps
Injectate volume 10 cc
Injectate temperature 0 °C or 24 °C
Cardiac output 2.5, 5, 10 liters per minute ± 7.5 %
Faulty-injectate curve Waveform for simulation available
Left-to-right-shunt curve Waveform for simulation available
Calibrated pulse 1.5 ° for 1 second
Connector Circular DIN 7 pin
Power-on default 5 liters per minute, 0 °C injectate, 37 °C blood temperature
Non-invasive blood pressure
Pressure units  
Manometer (Pressure meter)  
Pressure source  
NIBP simulations Pulse 2 mmHg max into 500 ml NIBP system
Volume of air moved 1.25 ml max
Simulations (systolic/diastolic [MAP]) Adult: 60/30 (40), 80/50 (60); 100/65 (77); 120/80 (93); 150/100 (117); and 200/150 (167) and 255/195 (215) Neonatal: 35/15 (22); 60/30 (40); 80/50 (60); 100/65 (77); 120/80 (93) and 150/100 Pressure variability: systolic and diastolic pressures are variable by 1 mmHg
Repeatability Within ± 2 mmHg (at maximum pulse size independent of device under test)
Synchronization: normal Sinus heart rates: 30 BPM to 240 BPM Maximum rate at 1 ml: 240 BPM achievable with pulses up to 1 ml Maximum rate at 1.25 ml: 180 BPM
Synchronization: arrhythmias Premature atrial contraction (PAC), premature ventricular contraction (PVC), atrial fibrillation, and missed beat
Leak test Target pressure 20 mmHg to 400 mmHg
Elapse time 0:30 min to 5:00 minutes: seconds in 30 second steps
Leakage rate 0 mmHg/minute to 200 mmHg/ minute
Pressure relief test range 100 to 400 mmHg
For full Fluke ProSim 8 Vital Signs Simulator product specifications, please click here: ProSim 8

Fluke Biomedical ProSim 8 Vital Signs Simulator

8-in-1 multifunction simulator tests ECG (including fetal ECG and arrhythmias), respiration, temperature, IBP, cardiac output, NIBP, SpO2, and Rainbow multi-wavelength waveforms

ProSim 8 patient simulator enables you to complete preventive maintenance testing with one device in less than five minutes, reducing average test time by up to 88%

Includes Carrying Case & Accessories As Listed

Fluke BioMedical ProSim 8 ProSim8 Vital Signs Simulator Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.

 

BC# 43431-L/T*

  • Fluke Biomedical ProSim 8 Vital Signs Simulator (BC# 43431-L/T*)
  • Fluke Biomedical ProSim Spot Sp02 Test Module
  • Fluke Biomedical Ansur Test Automation Software
  • AC Adapter/Charger
  • USB PC Interface Cable
  • User Manual On CD-ROM
  • Hard Carrying Case w/ Strap