Fluke BioMedical VT Plus HF Gas Flow Ventilator Analyzer

$4,499.00
Availability: In stock
SKU
MT25

Condition: Refurbished

Warranty: 1 Year Warranty

VT Plus HF Gas Flow Ventilator Analyzer

The VT PLUS HF is Fluke Biomedical’s premier general-purpose gas-flow analyzer. In addition, special display modes and bi-directional flow make it perfect for fully and efficiently testing both conventional mechanical ventilators and high-frequency ventilators. EC.6.20 now requires 100 % completion of scheduled life-support device preventive maintenance every year, and VT PLUS HF can help meet those requirements. Multiple special-function tests make troubleshooting quick and efficient.

 

VT PLUS HF has the capability to measure either high- or low-flow and pressure, replacing the need for gauges and flow meters. It measures 21 ventilator parameters and can display all of them on one screen. Results can be printed directly from the unit or from a PC with included Windowscompatible software. VT PLUS HF also has onboard graphing capability and shows the minimum, maximum, average, and absolute measurement for all parameters.

 

Learning to use the VT PLUS HF is simple. Technicians control the unit using the VT PLUS HF user-friendly command system, or, if they’re familiar with the RT-200, they can switch to a special control mode that uses RT-200-style commands.

 

VT PLUS HF can be operated with a variety of precision test lungs to ensure that ventilators are tested to manufacturers’ specifications and clinical expectations with a fully NIST-traceable testing system.

 

Features:

  • Bi-directional flow, pressure, volume, and oxygen concentration, and pressure measurements
  • Low- and high-pressure, and flow measurement capability
  • Special HF mode—up to 900 BPM (15 Hz)
  • RS-232 and printer ports
  • Included Windows-compatible graphics software
  • All 21ventilator parameters displayed at once on one screen
  • Operation by user-friendly VT PLUS HF command mode or special RT-200 command mode
  • Minimum, maximum, average, absolute, and graph for all parameters
  • Multiple special-function tests for efficient troubleshooting

Specifications;b

Power 100 V ac to 240 V ac, 50/60 Hz
Maximum over-voltage 264 V ac
Power consumption < 132 V A
Fuse rating 0.5 A, slow blow
Display 320 x 240 LCD with CFL backlight
Viewing area 10.1 cm x 8.2 cm (3 in x 4 in), blue on white background
Operational modes Manual mode for simple tests or troubleshooting; computer-control mode, using RS-232 serial port for special applications; use of VT PLUS HF with VT for Windows software for recording graphs and logging data to a computer
Output ports RS-232 serial port, and parallel-printer port
Oxygen measurement
Range 0 % to 100 %
Accuracy ± 2 % FSO
Resolution 0.1 % O2
Transducer location Internal
Gas
Compatibility Air, O2, CO2, N2, N2O, He, mixtures, or user-defined
Reference units ATP, STPD0, STPD21 and BTPS
Test parameters
Continuous flow Low flow: ± (2 % of reading and 1 % of range)
High flow: ± (2 % of reading and 1 % of range)
Volumetric flow Low-flow
Flow range: -25 lpm to 25 lpm
Accuracy: ± 2 % of reading or ± 1 % of range, whichever is greater
Frequency response: > 25 Hz or t10-90 < 40 ms, whichever is greater
Low-flow dropout: 0.01 lpm
Breath-detect threshold: 0.5 lpm
Maximum-flow rate: 50 lpm
Volume range: > ± 60 l
Sample rate: 100 Hz
Resolution: 0.01 lpm flow > 1 lpm; 0.001 lpm flow < 1 lpm
Dynamic resistance: < 2.5 cmH20 @ 5 lpm
Fittings: 15 mm OD, 1:40 conical male; 0.25 in NPT ID per ASTM F-1054 aluminum with black anodized finish
High-flow
Flow range: -300 lpm to 300 lpm
Accuracy: ± 2 % of reading or ± 2 % of range, whichever is greater
Frequency response: > 25 Hz
High-flow dropout: 25 lpm
Breath-detect threshold: 2 lpm
Maximum-flow rate: 500 lpm
Volume range: > ± 60 l
Sample rate: 100 Hz
Resolution: 0.01 lpm
Dynamic resistance: 2 cmH20 @ 60 lpm
Fittings: 22 mm OD, 1:40 conical male; 15 mm ID, 1:40 conical female per ASTM F-1054 aluminum with black anodized finish
High-pressure
Maximum applied pressure: 150 psi
Range: ± 100 psi
Accuracy: ± 1 % of reading or ± 0.3 psig, whichever is greater
Frequency response: > 10 Hz
Resolution: 0.1 psi
Sample rate: 100 Hz
Fittings: DISS connector, stainless steel
Airway-pressure
Maximum applied pressure: 20 psi
Range: ± 120 cmH2O
Accuracy: ± 0.75 % of reading or ± 0.5 cmH2O, whichever is greater
Frequency response: > 25 Hz or t10-90 < 40 ms, whichever is greater
Resolution: 0.1 cmH2O
Sample rate: 100 Hz
Fittings: Internally connected at the transducer distal end
Ventilator parameter 
Inspiratory and expiratory tidal volume RResolution: 0.1 ml
Range: As specified in high-flow/low-flow specification
Accuracy: As specified in high-flow/low-flow specification
Expirartory iniute volume/td> RResolution: 0.001 lpm
Range: 0 L to 60 L
Accuracy: ± 3 %
Breath rate/td> Resolution: 0.1 BPM
Range: 0.5 BPM to 150 BPM
Accuracy: ± 1 %
Inspitaory to expiratory time ratio (I:E Ratio)/td> Resolution: 0.01
Range: 1:200 to 200:1
Accuracy: ± 2 % or ± 0.1 s
Inspiratory time/td> Resolution: 0.01 s
Range: 0 s to 60 s
Accuracy: ± 1 % or ± 0.02 s
Expiratory time/td> Resolution: 0.01 s
Range: 0 s to 90 s
Accuracy: ± 1 % or ± 0.01 s
Peak inspiratory pressure/td> Resolution: 0.1 cmH20
Range: ± 120 cmH2O
Accuracy: ± 3 % or ± 1 cmH2O
Inspriatory pausre pressure/td> Resolution: 0.1 cmH20
Range: ± 120 cmH2O
Accuracy: ± 3 % or ± 1 cmH2O
Men airway pressure/td> Resolution: 0.1 cmH20
Range: ± 80 cmH2O
Accuracy: ± 3 % or ± 0.5 cmH2O
Positive end expiratory pressure (PEEP)/td> Resolution: 0.1 cmH20
Range: -5 cmH2O to 40 cmH2O
Accuracy: ± 3 % or ± 0.5 cmH2O
Lung compliance/td> Resolution: 0.1 ml/cmH20
Range: 0 ml/cmH2O to 150 ml/cmH2O
Accuracy: ± 5 % or ± 5 ml/cmH2O
Inspriatory hold time/td> Resolution: 0.01 s
Range: 0 s to 60 s
Accuracy: ± 1 % or ± 0.1 s
Expiratory hold time/td> Resolution: 0.01 s
Range: 0 s to 90 s
Accuracy: ± 1 % or ± 0.1 s
Peak expiratory flow/td> Resolution: 0.01 lpm
Range: 0 lpm to 300 lpm
Accuracy: ± 3 % or ± 2 lpm
Peak inspiratory flow/td> Resolution: 0.01 lpm
Range: 0 lpm to 300 lpm
Accuracy: ± 3 % or ± 2 lpm
Flow bias/td> Resolution: 0.01 lpm
Range: 0 lpm to 30 lpm
Accuracy: ± 2 % or ± 0.5 lpm
Operating environment conditions
Temperature range 10 °C to 40 °C
Ambient humidity 0 % to 80 % non-condensing to 31 °C, decreasing to 50 % at 40 °C
Barometric pressure 8 psig to 18 psig
Storage environment conditions
Temperature range -25 °C to -50 °C
Humidity 0 to 95 % non-condensing
Dimensions (WxDxH) 25.4 cm x 25.4 cm x 12.7 cm (10 in x 10 in x 5 in)
Weight 4.53 kg (10 lb)

 

 

 

For full Fluke VT Plus HF Gas Analyzer product specifications, please click here: Fluke VT Plus HF

Fluke VT Plus HF Gas Flow Analyzer

Special display modes and bi-directional flow make this perfect for fully and efficiently testing both conventional mechanical ventilators and high-frequency ventilators

Includes Accessories As Listed

Fluke BioMedical VT Plus HF VTPlusHF Gas Flow Analyzer Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.

 

BC# 40283-L/T*

  • Fluke VT Plus HF Gas Flow Analyzer (BC# 40283-L/T*)
  • Power Cord
  • RS232 PC Interface Cable
  • User Manual On CD-ROM
Request a Quote for

Have Questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

we are commited to win your business!