Fluke BioMedical VT Plus HF Gas Flow Ventilator Analyzer
Condition: Refurbished
Warranty: 1 Year Warranty
Shipping: USA & Worldwide
- Fluke VT Plus HF Gas Flow Analyzer (BC# 40283-L/T*)
- Power Cord
- RS232 PC Interface Cable
- User Manual On CD-ROM
VT Plus HF Gas Flow Ventilator Analyzer
The VT PLUS HF is Fluke Biomedical’s premier general-purpose gas-flow analyzer. In addition, special display modes and bi-directional flow make it perfect for fully and efficiently testing both conventional mechanical ventilators and high-frequency ventilators. EC.6.20 now requires 100 % completion of scheduled life-support device preventive maintenance every year, and VT PLUS HF can help meet those requirements. Multiple special-function tests make troubleshooting quick and efficient.
VT PLUS HF has the capability to measure either high- or low-flow and pressure, replacing the need for gauges and flow meters. It measures 21 ventilator parameters and can display all of them on one screen. Results can be printed directly from the unit or from a PC with included Windowscompatible software. VT PLUS HF also has onboard graphing capability and shows the minimum, maximum, average, and absolute measurement for all parameters.
Learning to use the VT PLUS HF is simple. Technicians control the unit using the VT PLUS HF user-friendly command system, or, if they’re familiar with the RT-200, they can switch to a special control mode that uses RT-200-style commands.
VT PLUS HF can be operated with a variety of precision test lungs to ensure that ventilators are tested to manufacturers’ specifications and clinical expectations with a fully NIST-traceable testing system.
Features:
- Bi-directional flow, pressure, volume, and oxygen concentration, and pressure measurements
- Low- and high-pressure, and flow measurement capability
- Special HF mode—up to 900 BPM (15 Hz)
- RS-232 and printer ports
- Included Windows-compatible graphics software
- All 21ventilator parameters displayed at once on one screen
- Operation by user-friendly VT PLUS HF command mode or special RT-200 command mode
- Minimum, maximum, average, absolute, and graph for all parameters
- Multiple special-function tests for efficient troubleshooting
Specifications;b
Power | 100 V ac to 240 V ac, 50/60 Hz |
Maximum over-voltage | 264 V ac |
Power consumption | < 132 V A |
Fuse rating | 0.5 A, slow blow |
Display | 320 x 240 LCD with CFL backlight |
Viewing area | 10.1 cm x 8.2 cm (3 in x 4 in), blue on white background |
Operational modes | Manual mode for simple tests or troubleshooting; computer-control mode, using RS-232 serial port for special applications; use of VT PLUS HF with VT for Windows software for recording graphs and logging data to a computer |
Output ports | RS-232 serial port, and parallel-printer port |
Oxygen measurement | |
Range | 0 % to 100 % |
Accuracy | ± 2 % FSO |
Resolution | 0.1 % O2 |
Transducer location | Internal |
Gas | |
Compatibility | Air, O2, CO2, N2, N2O, He, mixtures, or user-defined |
Reference units | ATP, STPD0, STPD21 and BTPS |
Test parameters | |
Continuous flow | Low flow: ± (2 % of reading and 1 % of range) High flow: ± (2 % of reading and 1 % of range) |
Volumetric flow | Low-flow Flow range: -25 lpm to 25 lpm Accuracy: ± 2 % of reading or ± 1 % of range, whichever is greater Frequency response: > 25 Hz or t10-90 < 40 ms, whichever is greater Low-flow dropout: 0.01 lpm Breath-detect threshold: 0.5 lpm Maximum-flow rate: 50 lpm Volume range: > ± 60 l Sample rate: 100 Hz Resolution: 0.01 lpm flow > 1 lpm; 0.001 lpm flow < 1 lpm Dynamic resistance: < 2.5 cmH20 @ 5 lpm Fittings: 15 mm OD, 1:40 conical male; 0.25 in NPT ID per ASTM F-1054 aluminum with black anodized finish High-flow Flow range: -300 lpm to 300 lpm Accuracy: ± 2 % of reading or ± 2 % of range, whichever is greater Frequency response: > 25 Hz High-flow dropout: 25 lpm Breath-detect threshold: 2 lpm Maximum-flow rate: 500 lpm Volume range: > ± 60 l Sample rate: 100 Hz Resolution: 0.01 lpm Dynamic resistance: 2 cmH20 @ 60 lpm Fittings: 22 mm OD, 1:40 conical male; 15 mm ID, 1:40 conical female per ASTM F-1054 aluminum with black anodized finish High-pressure Maximum applied pressure: 150 psi Range: ± 100 psi Accuracy: ± 1 % of reading or ± 0.3 psig, whichever is greater Frequency response: > 10 Hz Resolution: 0.1 psi Sample rate: 100 Hz Fittings: DISS connector, stainless steel Airway-pressure Maximum applied pressure: 20 psi Range: ± 120 cmH2O Accuracy: ± 0.75 % of reading or ± 0.5 cmH2O, whichever is greater Frequency response: > 25 Hz or t10-90 < 40 ms, whichever is greater Resolution: 0.1 cmH2O Sample rate: 100 Hz Fittings: Internally connected at the transducer distal end |
Ventilator parameter | |
Inspiratory and expiratory tidal volume | RResolution: 0.1 ml Range: As specified in high-flow/low-flow specification Accuracy: As specified in high-flow/low-flow specification |
Expirartory iniute volume/td> | RResolution: 0.001 lpm Range: 0 L to 60 L Accuracy: ± 3 % |
Breath rate/td> | Resolution: 0.1 BPM Range: 0.5 BPM to 150 BPM Accuracy: ± 1 % |
Inspitaory to expiratory time ratio (I:E Ratio)/td> | Resolution: 0.01 Range: 1:200 to 200:1 Accuracy: ± 2 % or ± 0.1 s |
Inspiratory time/td> | Resolution: 0.01 s Range: 0 s to 60 s Accuracy: ± 1 % or ± 0.02 s |
Expiratory time/td> | Resolution: 0.01 s Range: 0 s to 90 s Accuracy: ± 1 % or ± 0.01 s |
Peak inspiratory pressure/td> | Resolution: 0.1 cmH20 Range: ± 120 cmH2O Accuracy: ± 3 % or ± 1 cmH2O |
Inspriatory pausre pressure/td> | Resolution: 0.1 cmH20 Range: ± 120 cmH2O Accuracy: ± 3 % or ± 1 cmH2O |
Men airway pressure/td> | Resolution: 0.1 cmH20 Range: ± 80 cmH2O Accuracy: ± 3 % or ± 0.5 cmH2O |
Positive end expiratory pressure (PEEP)/td> | Resolution: 0.1 cmH20 Range: -5 cmH2O to 40 cmH2O Accuracy: ± 3 % or ± 0.5 cmH2O |
Lung compliance/td> | Resolution: 0.1 ml/cmH20 Range: 0 ml/cmH2O to 150 ml/cmH2O Accuracy: ± 5 % or ± 5 ml/cmH2O |
Inspriatory hold time/td> | Resolution: 0.01 s Range: 0 s to 60 s Accuracy: ± 1 % or ± 0.1 s |
Expiratory hold time/td> | Resolution: 0.01 s Range: 0 s to 90 s Accuracy: ± 1 % or ± 0.1 s |
Peak expiratory flow/td> | Resolution: 0.01 lpm Range: 0 lpm to 300 lpm Accuracy: ± 3 % or ± 2 lpm |
Peak inspiratory flow/td> | Resolution: 0.01 lpm Range: 0 lpm to 300 lpm Accuracy: ± 3 % or ± 2 lpm |
Flow bias/td> | Resolution: 0.01 lpm Range: 0 lpm to 30 lpm Accuracy: ± 2 % or ± 0.5 lpm |
Operating environment conditions | |
Temperature range | 10 °C to 40 °C |
Ambient humidity | 0 % to 80 % non-condensing to 31 °C, decreasing to 50 % at 40 °C |
Barometric pressure | 8 psig to 18 psig |
Storage environment conditions | |
Temperature range | -25 °C to -50 °C |
Humidity | 0 to 95 % non-condensing |
Dimensions (WxDxH) | 25.4 cm x 25.4 cm x 12.7 cm (10 in x 10 in x 5 in) |
Weight | 4.53 kg (10 lb) |
For full Fluke VT Plus HF Gas Analyzer product specifications, please click here: Fluke VT Plus HF