The i2000 is an AC current probe utilizing the Rogowski principle. It can be used to measure currents up to 2000A when used in conjunction with multimeters, recorders or data loggers. The flexible and lightweight measuring head allows quick and easy installation in hard to reach areas.
Three ranges: 20A, 200 A & 2000 A, 100 mV/A, 10 mV/A & 1 mV/A output
Ideal for current measurement on thick and hard to reach conductors without breaking the circuit
Maximum conductor diameter 178 mm
CAT III 600V safety rating
Specifications
Electrical specifications
Current ranges
20 A
200 A
200 A
AC rms
Output sensitivity (AC-coupled)
100 mV
10 mV
1 mV
1 mV per A
Load impedance
100 kO minimum
Accuracy
at +25 °C
± 1 % of range (45 Hz to 65 Hz)
Linearity
10 % to 100 % of range
± 0.2 % of reading
Noise
8 mV ac rms (30 A)
2 mV ac rms (300 A, 3000 A)
Frequency range
10 Hz to 20 kHz (-3 dB)
Phase error
< ± 1 º (45 Hz to 65 Hz)
± 10 º (at 20 kHz)
Position sensitivity
with cable > 25 mm from the coupling
± 2 % of range
External field
with cable > 200 mm from the head
± 1 % of range
Power supply
2 x AA/MN 1500/LR6 Alkaline, 200 hours, low battery indicator
Temperature Coefficient
± 0.08% of reading per°C
Working voltage
see Safety standards section
600 V AC rms or DC
General Specifications
Head Circumference
610 mm (24 in)
Head cable diameter
14.3 mm (0.56 in)
Head bend radius
38.1 mm (1.5 in)
Cable length (head to electronics)
2 m (6.56 ft)
Output connection
Molded 4 mm safety plugs (not integrated)
Operating temperature range
0 °C to +70 °C (32 °F to 158 °F) (head and electronics)
Storage temperature range
-40 °C to +105 °C (-40 °F to 221 °F) (head)
-20 °C to +85 °C (-4 °F to 185 °F) (electronics)
Operating humidity
15 % to 85 % (non-condensing)
Weight
180 g (6.34 oz) (head)
190 g (6.70 oz) (electronics)
Safety standards
EN 61010-1: 2001
EN 61010-2-032: 2002
EN 61010-031: 2002
600 Vrms, Category III, Pollution Degree 2
Use of the probe on uninsulated conductors is limited to 600 V ac rms or dc and frequencies below 1 kHz.
EMC standards
EN 61326: 1998 +A1, A2, & A3
Dimensions
HxWxD
116 mm x 68.5 mm x 30 mm (4.56 in x 2.69 in x 1.18 in)