LeCroy AP015 30 A 50 MHz AC Current Probe

LeCroy AP015 30 A 50 MHz AC Current Probe

LeCroy CP150 150 Amp 10 MHz AC/DC Current Probe CP-150

LeCroy CP150 150 Amp 10 MHz AC/DC Current Probe CP-150

LeCroy AP033 500 MHz Differential Active Probe

$1,895.00
Availability: In stock
SKU
11H75

Condition: Refurbished

Warranty: 1 Year Warranty

LeCroy AP033 Differential Active Probe

 

The AP033 is a wide band differential active probe. The probe features low noise, low input capacitance, high common mode rejection, and FET-buffered inputs in the probe head. User selectable attenuation and offset give the probe flexibility to measure a large range of signal amplitudes. Plug-on attenuator and AC coupling accessories further extend the application range. Interconnect accessories included allow connection to surface mount and through-hole components with minimal signal degradation. The input receptacles in the probe head are compatible with standard 0.025 in. (0.635 mm) square pins. This provides a convenient low cost method of creating device characterization test fixtures.

 

The probe is powered directly from a Teledyne LeCroy oscilloscope through the ProBus® interface. The ProBus interface also allows local control of the probe through the oscilloscope user interface and remote control through the interface buses, (GPIB, RS-232). The optional ADPPS power supply allows the AP033 to be used with other instruments such as spectrum analyzers, network analyzers, and oscilloscopes without ProBus interface.

 

Applications

The AP033 is ideal for acquiring high speed differential signals such as those found in disk drive read channels, differential LAN, video, etc. It can also be used with spectrum analyzers to acquire signals in some RF systems (for example, balanced IF mixers in hand held cellular telephones). The high impedance characteristics of both inputs allow the probe to be used as a FET probe to make single-ended measurements in digital systems without introducing a ground loop, as a conventional FET probe would.

 

 

Specifications:

Bandwidth 500 MHz
Input Configuration: True Differential (+ and – Inputs); with shield Ground connector
Effective Gain : X1, X10, ÷10, ÷1002
Input coupling: DC. AC Coupling obtained by installing AC Coupling Adapter.
Differential Mode Range: (10X Gain): ±40 mV (÷1 Attenuation) ±400 mV (÷10 Attenuation) ±4 V (÷100 Attenuation) (1X Gain): ±400 mV (÷1 Attenuation) ±4 V (÷10 Attenuation) ±40 V (÷100 Attenuation) Common Mode Range: ±4.2 V (÷1 Attenuation) ±42 V (÷10 Attenuation) ±42 V (÷100 Attenuation)
Maximum Input Voltage: ±42 V either input from ground
Electrical Characteristics
LF Gain Accuracy: 2% into 50 Ω load3 , measured at 1 kHz with 0 V offset
Common Mode Rejection Ratio (Probe head grounded, DC Coupled, ÷1 attenuation without external attenuator): 70 Hz ≥ 3160:1 (70 dB)
1 MHz ≥ 1000:1 (60 dB)
250 MHz ≥ 5:1 (14 dB)
For full LeCroy AP033 Digital Probe product specifications, please click here: AP033

LeCroy AP033 500 MHz Differential Active Probe

The AP033 is ideal for acquiring high speed differential signals such as those found in disk drive read channels, differential LAN, video, etc. It can also be used with spectrum analyzers to acquire signals in some RF systems

LeCroy AP033 AP-033 Differential Active Probe Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.

 

BC# 22919M-L/T*

  • LeCroy AP033 500 MHz Differential Active Probe (BC# 22919M-L/T*)
Request a Quote for

Have Questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

we are commited to win your business!