Megger S1-552/2 5 kV Insulation Resistance Tester
$2,329.00
Availability:
In stock
SKU
MG03
Condition: Refurbished / Calibrated
Warranty: 1 Year Warranty
Shipping: USA & Worldwide
- Megger S1-552/2 5 kV Insulation Resistance Tester (BC# 18836-L/T*)
- Megger T/V Test Set (1-Red, 1-Black, 1-Blue)
- Megger Quick Start Guide
- Megger Test Lead User Guide
- Attached Accessory Case
- Power Cord
- RS232 PC Interface Cable
- USB PC Interface Cable
- Integrated Hard Case
Request a Quote for
Megger S1-552/2 Insulation Resistance Tester
The new 5 & 10 kV insulation resistance testers from Megger are designed specifically to assist you with the testing and maintenance of high voltage electrical equipment. The case is incredibly rugged and easy to carry, being made of tough polypropylene and achieving an ingress protection rating of IP65. In addition, the model number is marked on both sides of the case for ease of identification in stores or vehicles.
The instruments have a large easy to read backlit LCD display making it equally suitable for use in both bright sunlight and poorly lit environments. Information displayed includes resistance, voltage, leakage current, capacitance, battery status and time constant. In addition, the elapsed time of the test is displayed constantly, removing the need for separate timers.
The instrument can test when being powered by the mains or its internal rechargeable battery, a great benefit when site conditions are unknown or long term testing is required.
The controls of the instrument are clear and unambiguous and a モquick startヤ guide is included in the lid of the instrument removing the need to carry bulky manuals under site conditions.
The instrument is fitted with a guard terminal to enhance accuracy. The guard test lead is included as standard with the instrument.
To further enhance the flexibility of the instruments both have the facility to set the test voltage in steps from 50 V to the maximum output voltage avoiding the expense of several insulation resistance testers to meet your application needs.
These IR instruments are designed to meet the highest safety standards and meet the requirements of EN61010. In addition, the instrument measures the voltage at the terminals and if this is above 50 V then the instrument will display the high voltage warning and inhibit testing. At the end of the test the instrument will automatically discharge the energy transferred to the equipment during the test phase.
The S1-552/2 is suited to routine testing and diagnostic regimes being pre-programmed with IR, PI, DAR, SV and DD profiles.
Features:
Specifications:
For full Megger S1-552/2 specifications, please click here: Megger S1-552/2
The new 5 & 10 kV insulation resistance testers from Megger are designed specifically to assist you with the testing and maintenance of high voltage electrical equipment. The case is incredibly rugged and easy to carry, being made of tough polypropylene and achieving an ingress protection rating of IP65. In addition, the model number is marked on both sides of the case for ease of identification in stores or vehicles.
The instruments have a large easy to read backlit LCD display making it equally suitable for use in both bright sunlight and poorly lit environments. Information displayed includes resistance, voltage, leakage current, capacitance, battery status and time constant. In addition, the elapsed time of the test is displayed constantly, removing the need for separate timers.
The instrument can test when being powered by the mains or its internal rechargeable battery, a great benefit when site conditions are unknown or long term testing is required.
The controls of the instrument are clear and unambiguous and a モquick startヤ guide is included in the lid of the instrument removing the need to carry bulky manuals under site conditions.
The instrument is fitted with a guard terminal to enhance accuracy. The guard test lead is included as standard with the instrument.
To further enhance the flexibility of the instruments both have the facility to set the test voltage in steps from 50 V to the maximum output voltage avoiding the expense of several insulation resistance testers to meet your application needs.
These IR instruments are designed to meet the highest safety standards and meet the requirements of EN61010. In addition, the instrument measures the voltage at the terminals and if this is above 50 V then the instrument will display the high voltage warning and inhibit testing. At the end of the test the instrument will automatically discharge the energy transferred to the equipment during the test phase.
The S1-552/2 is suited to routine testing and diagnostic regimes being pre-programmed with IR, PI, DAR, SV and DD profiles.
Features:
- CAT IV 600 V
- Mains or battery powered
- Digital/analogue backlit display
- Variable test voltage from 50 to 5000 or 10,000 V
- Automatic IR, PI, DAR, SV and DD tests
- Measures to 15 TO (5 kV)
- Charge current: 5 mA
- On board memory for results storage
Specifications:
Voltage input range | 85-265 V rms, 50/60Hz, 60 VA | ||||
S1-552/2 battery life | Typical capacity is 6 hours continuous testing at 5 kV with a 100 MO load | ||||
S1-552/2 test voltages | 50 V to 1 kV in 10 V steps, 1 kV to 5 kV in 25 V steps | ||||
S1-552/2 accuracy (23 °C, 5 kV) | |||||
5000 V | 2500 V | 1000 V | 500 V | 250 V | |
±5% to | 1 TO | 500 GO | 200 GO | 100 GO | 50 GO |
±20% to | 10 TO | 5 TO | 2 TO | 1 TO | 500 GO |
Guard | 2% error guarding 500 kO leakage with 100 MO load | ||||
Display range | Digital display (3 digits) 10 kO to 15 TO | ||||
Short circuit/charge current | 5 mA @ 5 kV | ||||
Capacitor charge time | <1.5 seconds per µF at 5mA to 5 kV | ||||
Capacitor discharge time | <120ms per µF to discharge from 5000 V to 50 V | ||||
Capacitance measurement (above 500 V) | 10 nF to 50 µF | ||||
Capacitance measurement accuracy (23 °C) | ±5% ±5 nF | ||||
Voltage output accuracy (0 °C to 30 °C) | +4%, -0%, ±10 V of nominal test voltage at 1 GO load | ||||
Current measurement range | 0.01 nA to 5 mA | ||||
Current measurement accuracy (23 °C) | ±5% ±0.2 nA at all voltages | ||||
Default voltmeter range: | 50-600 V a.c. or d.c. Accuracy (23 ºC) + or -2% + or -1 V | ||||
Interference rejection | 1 mA per 250 V up to a maximum of 3 mA | ||||
Timer range | Up to 99 minutes and 59 seconds from start of test 15 second minimum setting for test voltage =1000 V 30 second minimum setting for test voltage <1000 V |
||||
Memory capacity | 32 kB | ||||
Test regimes | Auto IR, PI, DAR, SV and DD. | ||||
Interface | RS232 and USB | ||||
Data store | Voltage, test time, leakage current, resistance, PI, DAR, DD, capacitance and time constant | ||||
Real time output | Serial, once per second of test voltage, current and resistance | ||||
Operating temperature | -20 °C to 50 °C | ||||
Storage temperature | -25 °C to 65 °C | ||||
Ingress protection (lid closed) | IP65 | ||||
Humidity | 90% RH non-condensing at 40 °C | ||||
Safety | Meets the requirements of IEC61010-1:2001 CATIV 600 V | ||||
EMC | Meets the requirements of IEC61326-1 | ||||
Dimensions | 305 x 194 x 360 mm (12.7 x 6 x 14.2 inches) | ||||
Weight | 7.1 kg (16lb) approx. |
For full Megger S1-552/2 specifications, please click here: Megger S1-552/2