Olympus Epoch 600 Ultrasonic Flaw Detector
Condition: Refurbished / Calibrated
Warranty: 1 Year Warranty
Shipping: USA & Worldwide
- Olympus Epoch 600 Ultrasonic Flaw Detector (BC# 42687-L/T*)
- AC Adapter/Charger
- Power Cord
- RS232 PC Interface Cable
- User Manual On CD-ROM
- Hard Carrying Case (G)
Olympus Epoch 600 Ultrasonic Flaw Detector
The EPOCH 600 Digital Ultrasonic Flaw Detector combines Olympus’ industry leading conventional flaw detection capabilities with the efficiency of a highly portable, intuitive instrument. The EPOCH 600 flaw detector’s blend of efficient menus and direct access keys allows you to take advantage of the highest quality flaw detection platform with exceptional ease of use.
The large, full VGA transflective display combined with our patent-pending digital high dynamic range receiver provides a stable, striking A-scan representation in any lighting condition. Available in two hardware configurations, the EPOCH 600 offers you the choice to adjust parameters with a knob or with a navigation area on the keypad. The EPOCH 600 is designed to meet the requirements of EN12668-1 and allows a full range of standard and optional flaw detection features. The rugged, ergonomic design allows use in nearly any inspection environment, while the flexible PerfectSquare™ pulser and digital filtering capabilities can tackle nearly any application.
Economy with Quality
The EPOCH 600 combines the efficiency of a basic flaw detector with the quality of Olympus’ most powerful fully digital ultrasonic instruments. A full range of dynamic, reliable inspection features are packaged with economy of size and ease of use. This optimal balance brings powerful flaw detection features within reach of any level of inspector.
Simple and Durable Operation
The EPOCH 600 design is focused on providing a very high level of flaw detection with the simplicity of a basic instrument. The EPOCH 600 is designed to be ergonomic, intuitive, and practical for both experienced and novice ultrasonic inspectors.
Intuitive User Interface
The EPOCH 600 user interface is based on the industry approved EPOCH 1000 Series imaging flaw detector. The EPOCH 600 combines a simple menu structure for instrument settings, calibration and software feature adjustment, with the EPOCH brand’s hallmark direct-access key approach for critical inspection functions such as gain and gate adjustment, screen freeze, and file save. Supported in multiple languages, the EPOCH 600 user interface is intuitive for any level of operator.
Portable Design for All Inspection Environments
The EPOCH 600 is designed for use in nearly any inspection environment, from bench top testing in a laboratory to extreme outdoor and hazardous conditions. Designed for IP rating in either knob (IP66) or navigation pad (IP67) configurations, and tested to very high environmental and reliability standards, the EPOCH 600 allows users in any inspection environment to feel confident in both the performance and durability of the instrument. The instrument is tested for vibration, shock, explosive atmosphere, and wide temperature ranges. And with a battery life to exceed 15 hours, the EPOCH 600 is the perfect solution for remote inspection locations.
Vibrant Full VGA Display
The EPOCH 600 features a full VGA (640 x 480 pixels) resolution display. The horizontal design of the EPOCH 600 optimizes the A-scan size and readability on this high quality display. Built with transflective technology, this VGA display provides excellent clarity in indoor, low lighting conditions using its powerful backlight, as well as in direct sunlight by using the ambient light as a pseudo-backlight.
Optimized Access to Powerful Features
The EPOCH 600 provides excellent quality ultrasonic performance. Based on the same digital architecture as the EPOCH XT, LTC and EPOCH 1000 Series, the EPOCH 600 provides flexible, powerful pulsing and receiving features to accommodate the needs of most flaw detection inspections.
Pulser/Receiver
The EPOCH 600 comes standards with powerful flaw detection capabilities, such as:
- PerfectSquare™ tunable square wave pulser
- Digital high dynamic range receiver
- Eight 100% digital filter sets
- Auto or manually adjustable PRF from 10 Hz to 2000 Hz
- Pulser voltage from 100 V to 400 V
- Amplitude resolution to ± 0.25%
- Five customizable digital measurements
Features:
- Designed to meet the requirements of EN12668-1
- PerfectSquare™ tunable square wave pulser
- Digital high dynamic range receiver
- Eight digital filters for enhanced signal-to-noise ratio
- 2 kHz PRF for rapid scanning
- Knob or navigation pad adjustment configurations
- Large, full VGA sunlight readable display
- Long battery life, supporting lithium-ion or alkaline batteries
- Standard dynamic DAC/TVG and onboard DGS/AVG
- 2 GB MicroSD memory card for data transfers and storage
- USB On-The-Go (OTG) for PC communication
- Alarm and VGA outputs
Specifications:
General | |
Overall dimensions (W x H x D) | 236 mm x 167 mm x 70 mm (9.3 in. x 6.57 in. x 2.76 in.) |
Weight | 1.68 kg (3.72 lb), including lithium-ion battery |
Languages | English, Spanish, French, German, Japanese, Chinese, Portuguese |
Transducer | BNC or Number 1 LEMO |
Data storage | 50,000 IDs onboard, removable 2 GB MicroSD card (standard) |
Battery type/life | Single lithium-ion rechargeable standard/12 h to 13 h (lithium-ion), 3 h (alkaline) |
Power requirements | AC Mains: 100 VAC to120 VAC, 200 VAC to240 VAC, 50 Hz to 60 Hz |
Display type | Full VGA (640 x 480 pixels) transflective color LCD, 60 Hz update rate |
Display dimensions (W x H, Diag.) | 117 mm x 89 mm, 146 mm (4.62 in. x 3.49 in., 5.76 in.) |
Pulser | |
Pulser | Tunable Square Wave |
PRF | 10 Hz to 2000 Hz in 10 Hz increments |
Energy settings | 100 V, 200 V, 300 V or 400 V |
Pulse width | Adjustable from 25 ns to 5,000 ns (0.1 MHz) with PerfectSquare™ technology |
Damping | 50, 100, 200, 400 Ω |
Gates | |
Measurement gates | 2 fully independent gates for amplitude and TOF measurements |
Gate start | Variable over entire displayed range |
Gate width | Variable from 2 to 95% full screen height |
Alarms | Positive and Negative Threshold, Minimum Depth (Gate 1 and Gate 2) |
Receiver | |
Gain | 0 to 110 dB |
Maximum input signal | 20 V p-p |
Receiver input impedance | 400 Ω ± 5% |
Receiver bandwidth | 0.2 MHz to 26.5 MHz at –3 dB |
Digital filter settings | Eight digital filter sets standard (0.2-10 MHz, 2.0- 21.5 MHz, 8.0-26.5 MHz, Battery type/life Single lithium-ion rechargeable standard/12 h to 13 h (lithium-ion), 3 h (alkaline) 0.5-4 MHz, 0.2-1.2 MHz, 1.5-8.5 MHz, 5-15 MHz, DC-10 MHz |
Rectification | Full-wave, Positive Half-wave, Negative Halfwave, RF |
System linearity | Horizontal : ± 0.2% FSW |
Vertical linearity | 0.25% FSH, amplifier accuracy ± 1dB |
Reject | 0 to 80% FSH with Visual Warning |
Amplitude measurement | 0 to 110% full screen height with 0.25% resolution |
Measurement rate | Equivalent to PRF in all modes |
Measurements | |
Measurement display locations | 5 locations available (manual or auto selection) |
Gate 1 | Thickness, Soundpath, Projection, Depth, Amplitude, Time-Of-Flight, Gates Min./Max. Depth, Min./Max. Amplitude |
Gate 2 | Same as Gate 1 |
Echo-to-Echo | Same as Gate 2-Gate 1 |
Other measurements | Overshoot (dB) value for DGS/AVG, ERS (equivalent reflector size) for DGS/AVG, AWS D1.1,D1.5 rating (D), Reject Value |
DAC.TVG | Standard |
DAC points | Up to 50 points, 110 dB dynamic range |
Special DAC modes | Custom DAC (up to 6 curves) |
Curved surface connection | Standard OD or Bar correction for Angle Beam measurements |