RAE Systems UltraRAE 3000 PGM7360 VOC Benzene Gas Detector
$5,809.00
Availability:
In stock
SKU
RS13
Condition: Refurbished
Warranty: 1 Year Warranty
Shipping: USA & Worldwide
- RAE Systems UltraRAE 3000 PGM7360 VOC & Benzene Gas Detector (BC# 42247-L/T*)
- 9.8eV Lamp
- RAE Travel Charger 059-3014-000
- RAE Flexi-I-Probe Assembly
- AC Adapter/Charger
- User Manual On CD-ROM
- Carrying Case (G)
Request a Quote for
RAE Systems UltraRAE 3000 VOC Monitor
The UltraRAE 3000's Photoionization Detector’s (PID) extended range of 0.05 to 10,000 ppm in VOC mode and 50 ppb to 200 ppm in benzene specific mode makes it an ideal instrument for applications, from entry pre-screening during refinery and plant maintenance to hazardous material response, marine spill response and refinery down-stream monitoring.
Features:
- Proven PID technology
- 3-second response time
- Extended range to 10,000 ppm
- Humidity sensor
- Automatic temperature-controlled sampling time calculation
- Highly specific readings
- New sampling probe design
- Versatile VOC or Benzene Specific modes
- Built-in Bluetooth
- Integrated RAE Systems Correction Factors list for more than 200 compounds
- Multi-language support
Specifications:
Size | 10" L x 3.0" W x 2.5" H (25.5 cm x 7.6 cm x 6.4 cm) |
Weight | 26 oz (738 g) |
Sensors | Photoionization sensor with standard 9.8 eV or optional 10.6 eV or 11.7 eV lamps |
Battery | - Rechargeable, external field-replaceable Lithium-Ion battery pack - Alkaline battery adapter |
Operating Hours | 16 hours of operation |
Keypad | 1 operation and 2 programming keys, 1 flashlight on/off button |
Direct Readout | - VOCs as ppm by volume - High and low values - STEL and TWA - Battery and shutdown voltage - Date, time, temperature |
Alarms | - 95dB buzzer (at 12”/30 cm) and flashing red LED to indicate exceeded preset limits - High: 3 beeps and flashes per second - Low: 2 beeps and flashes per second - STEL and TWA: 1 beep and flash per second - Alarms latching with manual override or automatic reset - Additional diagnostic alarm and display message for low battery and pump sta |
EMI/RFI | Highly resistant to EMI/RFI. Compliant with EMC directive (2004/108/EC); R & TTE directive (1999/5/EC) |
IP Rating | IP65 unit running |
Datalogging | Standard 6 months at one-minute intervals |
Calibration | 2-point or 3-point calibration for zero and span. Calibration memory for 8 calibration gases, alarm limits, span values and calibration dates |
Sampling Pump | - Internal, integrated flow rate at 400 cc/mn - Sample from 100’ (30m) horizontally and vertically |
Low Flow Alarm | - Auto pump shutoff at low-flow condition |
Communication | - Download data and upload instrument setup from PC through charging cradle - Wireless data transmission through built-in Bluetooth™ |
Wireless Network | ProRAE Guardian Real-Time Wireless Safety |
Wireless Emergency | ISM license-free bands |
Wireless Range | UltraRAE 3000 to RAELink3 or RAELink3 Z1 modems ~ 33 feet (10 meters) |
Hazardous Area Approval | - US and Canada: R , Classified as Intrinsically Safe for use in Class I, Division 1 Groups A, B, C, D. - Europe: ATEX II 2G EEx ia IIC T4 - IECEx CSA 10.0005 Ex ia IIC/IIB GbT4 |
Temperature | -4° to 122° F (-20° to 50° C) |
Humidity | 0% to 95% relative humidity (non-condensing) |
Attachments | Durable bright red rubber boot |