Rohde & Schwarz SGS100A 6GHz RF Source w/ Options B1 B26 B106 SGS-100

$7,799.00
Availability: Out of stock
SKU
9S95

Condition: Refurbished / Calibrated

Warranty: 1 Year Warranty

Shipping: USA & Worldwide

Request a Quote for

Have questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

We are commited to win your business!

  • Rohde & Schwarz SGS100A RF Source: (BC# -L/T*)
    • Installed Option:
      • B106: 1 MHz to 6 GHz, CW (no modulation)
      • B1: Reference Oscillator
      • B26: Step Attenuator
  • Power Cord
  • User Manual on CD

Rohde & Schwarz SGS100A RF Source

R&S®SGMA Product Family At a glance

 

The R&S®SGS100A is an RF source designed to meet the requirements of automated test systems. It is available as a CW source or as a vector signal generator with an integrated I/Q modulator. With its frequency range of up to 12.75 GHz, the vector signal generator version covers the essential digital signals. The CW version can be used as a flexible local oscillator as well as for interference testing against mobile radio standards. The R&S®SGU100A upconverter extends the frequency range to 40 GHz.

The very compact dimensions of only one height unit and ½ 19" rack width and the extremely fast frequency and level setting times are two features that demonstrate how consistently the R&S®SGS100A and the R&S®SGU100A have been designed for use in systems. They can also be used in all applications that require either multiple RF sig-nal sources, such as beamforming applications, or where an extremely compact RF generator is needed, e.g. an operating source in customer systems.

Though compact, the R&S®SGS100A together with the R&S®SGU100A provides uncompromising signal purity and level accuracy. It offers performance usually available only from high-end instruments, including high output power, level repeatability and modulation bandwidth. During development, particular emphasis was placed on the total cost of ownership. A long calibration interval, excel-lent serviceability and options for expansion maximize the availability of the RF source.

Since the RF source is typically remote controlled, the front panels of the R&S®SGS100A and the R&S®SGU100A have a minimalist design. However, they offer status LEDs as well as all of the keys necessary for controlling generator operation. When used as a benchtop instrument, the R&S®SGS100A and the R&S®SGU100A are operated manually using the R&S®SGMA-GUI software that comes with the instrument.

 

Key facts

Smallest fully integrated vector signal generator on the market, space-saving design for system integration

Enables high throughput due to very short frequency and level setting times of typ. 280 μs via PCIe interface

Excellent RF performance in a compact format

Maximum output level of typ. +22 dBm (for the R&S®SGS100A) for compensating losses in the setup

Closed ALC loop for CW and I/Q modes for highest level repeatability

Wear-free electronic attenuator for high reliability up to 12.75 GHz

Cost-efficient and compact frequency extension to 40 GHz with the R&S®SGU100A upconverter

Low initial costs and low cost of ownership

 

Benefits and key features

Smallest integrated signal generator: 1 HU, ½ 19" for minimized required space in the test system

Very fast setting times via PCIe interface for high throughput

High maximum level for compensating losses in the setup

Maximum level of typ. +22 dBm up to 12 GHz (with R&S®SGS-B26 electronic step attenuator option)

Maximum level of typ. +17 dBm up to 40 GHz (with the combination of the R&S®SGS100A and the R&S®SGU100A)

Optional wear-free electronic attenuator up to 12.75 GHz for reliable operation

 

High performance in the right package

Very high level linearity and repeatability thanks to a closed ALC loop with I/Q modulation up to 40 GHz for constant test conditions

Low SSB phase noise of typ. –133 dBc (20 kHz carrier offset, f = 1 GHz, 1 Hz measurement bandwidth)

Optional high-stability reference oscillator for easy integration of a standard into the test system

High modulation bandwidth of 2 GHz above 12 GHz carrier frequency with the R&S®SGU100A

 

Minimized total cost of ownership

Attractive initial cost

Straight forward modular design for short repair times, minimizing test system downtime

Long calibration interval for minimized service costs

 

Connectivity

 Simple integration into a test system

Multiple remote control options

R&S®SGMA-GUI PC software

 

Frequency extension to 40 GHz with the R&S®SGU100A upconverter

Two instruments into one: one RF output for the entire frequency range

Seamless integration into existing user interfaces❙Lowest form factor and power consumption on the market

High performance up to microwave frequencies

 

Specifications in brief  
   
Frequency  
Frequency Range  
With the R&S®SGS-B106 option 1 MHz to 6 GHz
With the R&S®SGS-B106 and R&S®SGS-B112 options 1 MHz to 12.75 GHz
With the R&S®SGS-B106 and R&S®SGS-B112 options plus the R&S®SGU100A with the R&S®SGU-B120 option 10 MHz to 20 GHz
With the R&S®SGS-B106 and R&S®SGS-B112 options plus the R&S®SGU100A with the R&S®SGU-B120 option and R&S®SGU-B140 option  10 MHz to 40 GHz
   
With the R&S®SGS-B106V option 80 MHz to 6 GHz
   
With the R&S®SGS-B106V and R&S®SGS-B112V options 80 MHz to 12.75 GHz
   
With the R&S®SGS-B106V and R&S®SGS-B112V options plus the R&S®SGU100A with the R&S®SGU-B120V option 80 MHz to 20 GHz
   
With the R&S®SGS-B106V and R&S®SGS B112V options plus the R&S®SGU100A with the R&S®SGU-B120V option and the R&S®SGU-B140V 80 MHz to 40 GHz
   
   
Setting time  
   
With PCIe remote control  
   
R&S®SGS100A alone < 500 μs
   
R&S®SGS100A together with R&S®SGU100A < 2 ms
   
   
Input frequency for external reference  
   
Into R&S®SGS100A 10 MHz, 100 MHz, 1000 MHz
   
Level
 
Specified level range  
   
For the R&S®SGS100A –10 dBm to +15 dBm (PEP) 1)
   
With the R&S®SGS-B26 option –120 dBm to +15 dBm (PEP) 1)
   
For the R&S®SGS100A together with the R&S®SGU100A  
   
50 MHz < f ≤ 12 GHz (bypass mode) –10 dBm to +15 dBm (PEP) 1)
   
12 GHz < f ≤ 20 GHz –10 dBm to +17 dBm (PEP) 1)
   
12 GHz < f ≤ 40 GHz                                                                         (with R&S®SGU-B140/B140V) –10 dBm to +15 dBm (PEP) 1)
   
Level error  
   
Setting Characteristic: AUTO; temperature range from +18 °C to +33 °C  
1 MHz ≤ f ≤ 3 GHz (R&S®SGS100A alone) < 0.5 dB
   
3 GHz < f ≤ 12.75 GHz (R&S®SGS100A alone) < 0.9 dB
   
12 GHz < f ≤ 40 GHz (R&S®SGS100A together with R&S®SGU100A), level > –30 dBm < 0.9 dB
   
Output impedance VSWR in 50 Ω system  
   
In full frequency range, with the R&S®SGS-B26 option (R&S®SGS100A alone < 1.8
   
In full frequency range (R&S®SGS100A together with R&S®SGU100A) < 1.7 (meas.)
   
   
Setting time  
   
With PCIe remote control, Setting Characteristic: AUTO  
   
R&S®SGS100A alone < 500 μs
   
R&S®SGS100A together with R&S®SGU100A < 2 ms (without switching of the mechanical attenuator)
   
R&S®SGS100A together with R&S®SGU100A plus R&S®SGU-B26 < 25 ms (with switching of the mechanical attenuator)
   
   
Spectral purity  
   
Harmonics level ≤ 8 dBm, CW, I/Q wideband OFF                                      < –30 dBc
   
Nonharmonics level > –10 dBm 2), offset > 10 kHz from carrier, f ≤ 1.5 GHz (for R&S®SGS100A)                                                                       < –76 dBc
   
Subharmonics level > –10 dBm 2), f ≤ 3 GHz                                                    < –76 dBc
   
Wideband noise  
   
1 MHz ≤ f ≤ 6 GHz and 10 MHz carrier offset, 6 GHz < f ≤ 12.75 GHz and 30 MHz carrier offset, auto mode, level > 5 dBm, 1 Hz measurement bandwidth, CW  
   
R&S®SGS100A alone < –145 dBc
   
R&S®SGS100A together with R&S®SGU100A  
   
CW, f ≤ 12 GHz  < –140 dBc (meas.)
   
CW, f > 12 GHz  < –142 dBc
   
CW, f > 19.5 GHz < -139 dBc (typ.)
   
   
SSB phase noise  
   
20 kHz carrier offset, 1 Hz measurement bandwidth  
   

   

F = 1 GHz< –126 dBc;  –130 dBc (typ.)
   
f = 10 GHz< –106 dBc;  –110 dBc (typ.)
   
   
I/Q modulation  
   
I/Q modulator bandwidth (RF)  
   
100 MHz < f ≤ 2.5 GHz, I/Q wideband  40 % of carrier frequency
   
2.5 GHz < f ≤ 12.25 GHz, I/Q wideband 1 GHz
   
F > 12 GHz (R&S®SGS100A together with R&S®SGU100A) 2 GHz
   
   
Error vector  
   
Measured with 16QAM, filter root cosine α = 0.5, 10 kHz symbol rate  
   
F > 80 MHz, RMS, 10 kHz symbol rate(R&S®SGS100A alone) < (0.4 % + 0.2 % × f/GHz)
   
F > 12 GHz, RMS, 10 MHz symbol rate(R&S®SGS100A together with R&S®SGU100A) < (2 % + 0.04 % × f/GHz)
   
ACLR  
   
WCDMA 3GPP FDD, TM 1/64, frequency range from 1800 MHz to 2200 MHz (R&S®SGS100A alone > 67 dB; 69.5 dB (meas.)
   
Baseband bypass mode  
   
Frequency range 1 MHz ≤ f ≤ 80 MHz
   
Specified level range  
   
R&S®SGS100A alone –5 dBm to +15 dBm
   
With the R&S®SGS-B26 option –120 dBm to +15 dBm
   
Remote control  
   
Using Rohde & Schwarz instrument driver PCIe (single lane)
   
Using SCPI 1999.5 or compatible command sets Ethernet (TCP/IP) 10/100/1000BaseT
   
Using SCPI 1999.5 or compatible command sets USB 2.0
   
   
General data  
   
Power consumption  
   
R&S®SGS100A alone 70 W (meas.)
   
R&S®SGU100A alone 40 W (meas.)
   
Dimensions W × H × D, per unit                                                                                     250 mm × 52.5 mm × 401 mm
(9.84 in × 2.07 in × 15.79 in)
1 HU, ½ 19" rack width
   
Weight when fully equipped 4.0 kg (8.82 lb

 

For full Rohde & Schwarz SGS100A RF Source product specifications, please click here: SGS100A