Tektronix 370 High Resolution Programmable Curve Tracer w/ Calibration Cert!
Condition: Refurbished / Calibrated
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- Tektronix 370 Programmable Curve Tracer (BC# 40821-L/T*)
- Tektronix A1009 EETS Module
- Tektronix A1007 Transistors Module
- Tektronix 012-0163-00 Module
- Tektronix A1002 IN-Line-Lead Module
- Calibration Certificate
Tektronix 370 Curve Tracer
Description
The 370 is a high-performance, GPIB-programable digital-storage curve tracer that provides static and dynamic semiconductor device measurements. This versatile instrument stimulates, measures, and displays the semiconductor characteristics of a variety of two, three, and four-terminal devices/ including bipolar transistors, field effect transistors, silicon-controlled rectifiers, diodes, thyristors, optoisolators, wafers, integrated circuits, etc. A variety of measurements can be performed using either grounded-emitter or grounded-base configurations.
The collector supply produces ac, rectified ac, or dc voltages ranging from 0 to ±2000 volts. This high voltage, combined with a current sensitivity of 100 pA/div, permits extended breakdown measurements on a device under test. A step generator produces voltage or current steps of either polarity for application to the base or emitter terminal. The step generator may also be operated in a pulsed mode to reduce DUT power dissipation.
In addition to conventional curve tracer performance, the 370 includes the following features:
1. Digital storage capability that allows bright and stable display and useful cursor measurements. The 370 can store up to 16 families of characteristic curves in a bubble memory cassette, display them on the crt, and send them for data processing via the GPIB. The bubble memory also provieds non-volatile storage for up to 16 complete front-panel setups.
2. Tow extended aquisition modes, called Averaging and envelope. Averaging reduces display noise in high sensitivity ranges. Envelope mode displays the maximum and minimum vertical or horizontal excursion of each curve, which is useful for detecting long-term variations such as thermal drift.
3. GPIB command-controllable front-panel functions. Nearly all are remotely controllable. (Exceptions are those controls intended only for manual operation, such as INTESITY, FOCUS, COLLECTOR SUPPLY HIGH-LOW control, etc. Also, curve data can be sent to or received from an external controller through the GPIB.
4. The CENTRONICS-compatible plotter interface permits sending displayed curve data and digital on-screen readouts to a digital plotter without an external controller.
5. Other features include an auxiliary voltage supply, cursor measurement readout, and diagnostic routines.
Installation
Initial Inspection
This instrument was thoroughly inspected for mechanical and electrical defects before shipment. It should be free of mars or scratches and meet or exceed all electrical specifications. To confirm this, inspect the instrument for physical damage incurred in transit and test the electrical performance by following the First Time Operation instructions in Section 3, Operating Instructions. For a complete verification of instrument performance, refer a qualified service technician to the performance check section of the service manual. If a discrepancy is found, contact your local Tektronix Field Office or representative.
Power Source Information
This instrument operates from a power source having a neutral at or near ground (earth) potential. It is not intended for operation from two phases of a multi-phase system, nor across legs of a single-phase, three wire system. This instrument can be operated from either a 115-Volt or 230-Volt nominal supply source, 48 to 66 Hz. Table 1-1 is a listing of the line voltage ranges, line frequency range, and power consumption.
Table 1-1 Line Voltage Ranges |
||
Setting of the RANGE switch | Setting of the NOMINAL switch | |
115VAC | 230VAC | |
HIGH | 107VAC to 132VAC | 214VAC to 250VAC |
LOW | 90VAC to 110VAC | 180VAC to 220VAC |
Line Requency Range | 48 TO 66 Hz | |
Power Consumption Max. Typical |
- 400W, 3.5A at 132V 60Hz 120W, 1.3 A at 115V 50Hz |
Electrical Specifications |
||
Characteristic |
Performance Requirement |
Operation Information |
DC Mode Ripple | 2% or less of voltage or 0.1% or less of full-range voltage. |
AC p-p open circuit Measurement |
Max Peak Volts LOW range HIGH range |
- 16V, 80V, 400V 2000V |
|
Voltage Accuracy | Peak open circuit voltage on all ranges withing +10, -0% |
at MAX PEAK POWER of 50 WATTS |
Range | 16V 80V 400V 2000V | - |
Max Peak Current | 10V 2A 0.4A 0.1A | |
Minimum Series Resistance (ohms) |
0.26 6.4 160 20K | |
Maximum Series Resistance (ohms) |
800 20K 500K 12.5M | |
Series Resistance Available (ohms, ±5% or ±0.1 ohm) |
0.26 1.3 6.4 32 160 800 4K 20K 100K 500K 2.5M 12.5M |
|
Peak Power Watts LOW range (16, 80 and 400V) |
220W 50W 10W 2W 0.4W 0.08W |
Derived from nominal peak open-circuit collector Vol- tages and nominal series resistance values |
HIGH range (200V) |
50W 10W 2W 0.4W 0.08W |
- |
Variable Collector Supply |
0 to 100.0% | Uncalibrated variable collector supply amplitude control from 0 to 100% in 0.1% increments. |
Resolution % Indicator (5LEDs) |
- | Indicates approximate % of MAX PEAK VOLTS |
Safety Interlocks | - | Applies to all ranges 16, 80, 400, 200V). The protective cover must be in place over test terminals and lid shut before voltage can be applied to the collector terminals. |
Collector Supply Disabled (LED) |
- | Amber light on indicates interlock is open. |
Warning Indicator | - | Red light on Indicates dangerous voltage maybe applied to collector terminal. |
Limiter Indicator | - | Indicates teh internal sensing circuit automatic protection is operating. |
Looping compensation | - | Cancels stray capacitance between collector terminal and ground. |
Accuracy (Current or votage steps incuding Offset) |
- | - |
Incremental | 1.5% | - |
Absolute | Less than 1.5% x total output + 3% x AMPLITUDE setting + 1 mV or 1 nA. (less than 1.5% of total output +10% of STEP/OFFSET setting +1mV or 1 nA with STEP Multi. 1X enabled.) |
- |
Offset Control Range Resolution |
Variable from -10 to +10 times STEP AMPLITUDE. STEP/OFFSET AMPLITUDE setting X1% |
- |
Current Mode Amplitude Range |
50nA to 200 mA in 1-2-5 sequence of 21 steps. |
Selected by STEP/OFFSET Amplitude |
Maximum Current | 20 Times STEP AMPLITUDE except 10X STEP AMPLITUDE when control is set to 200mA |
- |
Maximum Voltage | At least 10V | - |
Maximum Opposing Offset Current |
Ten times STEP AMPLITUDE | - |
Maximum Opposing Volts |
Less than 7V. | - |
* Check datasheet for complete specifications |
For full Tektronix 370 Curve Tracer product specifications, please click here: Tektronix 370