Tektronix TekScope THS710A 60 MHz Dual-Channel Oscilloscope 250MS/s
Condition: Refurbished
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- Tektronix TekScope THS710A 60MHz Oscilloscope 250MS/s (BC#-L/T*)
- (2) Agilent N2862B Voltage Probe Sets
- (2) Agilent Hard Point Test Leads
- AC Adapter/Charger
- Power Cord
- User Manual On CD-ROM
- Carrying Case (G)
The Tektronix THS710A oscilloscope is a graph-displaying device – it draws a graph of an electrical signal. In most applications, the graph shows how signals change over time: the vertical (Y) axis represents voltage and the horizontal (X) axis represents time. The intensity or brightness of the display is sometimes called the Z axis.
The Tektronix THS710A oscilloscope's simple graph can tell you many things about a signal, such as: the time and voltage values of a signal, the frequency of an oscillating signal, the “moving parts” of a circuit represented by the signal, the frequency with which a particular portion of the signal is occurring relative to, other portions, whether or not a malfunctioning component is distorting the signal, how much of a signal is direct current (DC) or alternating current (AC) and how much of the signal is noise and whether the noise is changing with time.
Product Description
The THS710A TekScope instruments combine a two-channel oscilloscope and a digital multimeter (DMM) in a rugged, handheld package.
General Features:
- Battery power or external power
- High-resolution, high-contrast display with temperature compensation for clear visibility over a wide temperature range
- Onboard waveform, data, and setup storage
- RS-232 communication port to load setups, download waveforms, and make hard copies
- Fully programmable through the RS-232 communication port
Oscilloscope Features:
The TekScope instrument is a powerful, two-channel oscilloscope with the following features:
- Autoranging for quick setup and hands-free operation
- 60 MHz THS710A bandwidth with selectable 20 MHz bandwidth limit
- 250 MS/s THS710A sample rate and 2,500 point record length
- Separate digitizers for each channel (both channels always acquire simultaneously)
- Waveform averaging and enveloping with hardware peak detection
- Digital Real Time digitizing (up to five-times oversampling), sin(x)/x interpolation, and peak-detect acquisition to limit the possibility of aliasing
- Independently isolated channels to allow improved safety for measurements to 1000 VRMS while floating up to 600 VRMS using P5102 probes
- Cursors and 21 continuously updated, automatic measurements
- Simultaneous oscilloscope and meter operation on the same or on separate signals
THS710A Specifications | ||
Acquisition Modes | Sample (Normal), Peak detect, Envelope, and Average | |
Acquisition Rate, typical | Up to 25 waveforms per second (2 channels, sample acquisition mode, MAG on, no measurements) | |
Single Sequence | ||
Acquisition Mode | Sample, Peak Detect Average, Envelope |
|
Acquisition Stops After | Single acquisition, one or two channels simultaneouslyN acquisitions, one or two channels simultaneously, N is settable from 2 to 256 or ∞ | |
Inputs | ||
Input Coupling | DC, AC, or GND | |
Input Impedance, DC Coupled | 1 M ±1% in parallel with 25 pF ±2 pF | |
Maximum Voltage Between Signal and Common at Input BNC | Overvoltage Category | Maximum Voltage |
CAT II Environment (refer to page A–19) | 300 VRMS | |
CAT III Environment (refer to page A–19) | 150 VRMS | |
For steady-state sinusoidal waveforms, derate at 20 dB/decade above 100 kHz to 13 Vpk at 3 MHz and above. Also, refer to Overvoltage Category description on page A–19. | ||
Maximum Voltage Between Common and Earth Ground at BNC | 600 VRMS (CAT II) or 300 VRMS (CAT III), using rated connectors or accessories 30 VRMS, 42.4 Vpk, using noninsulated connectors or accessories | |
Maximum Voltage, Channel-to-Channel Commons | 30 VRMS, 42.4 Vpk, using noninsulated connectors or accessories | |
Channel-to-Channel Common Mode Rejection, typical | 100:1 at frequencies ≤50 MHz, measured on MATH Ch1 – Ch2 waveform, with test signal applied between signal and common of both channels, and with the same VOLTS/DIV and coupling settings on each channe | |
Channel-to-Channel Crosstalk, typical | ≥100:1 at 50 MHz, measured on one channel, with test signal applied between signal and common of the other channel, and with the same VOLTS/DIV and coupling settings on each channel | |
Common to Chassis Capacitance, typical | 55 pF | |
Vertical | ||
Number of Channels | 2 | |
Digitizers | 8 bit resolution, separate digitizers for each channel sample simultaneously | |
VOLTS/DIV Range | 5 mV/div to 50 V/div at input BNC | |
Polarity | Normal and Invert | |
Position Range | ±10 divisions | |
Analog Bandwidth at BNC, DC Coupled (typical at 5 mV/div; guaranteed at all other settings) | 60 MHz at input BNC | |
Peak Detect or Envelope Bandwidth, typical (25 MS/s or slower) | 50 MHz | |
Analog Bandwidth Limit, typical | Selectable between 20 MHz or full | |
Lower Frequency Limit, AC Coupled, typical | ≤10 Hz at BNC, reduced by a factor of ten when using a 10X passive probe | |
Rise Time at BNC, typical | 5.8 ns | |
Peak Detect or Envelope Pulse Response, typical | Captures 50% or greater amplitude of pulses ≥ 8 ns wide (≥ 20 ns wide at 500 ns/div) | |
DC Gain Accuracy | ±2% for Sample or Average acquisition mode | |
Position Accuracy | ±[0.4% × (position × volts/div) + (0.1 div × volts/div)] | |
Oscilloscope | ||
Oscilloscope Sample Rate Range | 5 S/s to 250 MS/s, in a 1.25, 2.5, 5 sequence | |
Record Length | 2500 samples for each channel | |
SEC/DIV Range (including MAG) | 10 ns/div to 50 s/div | |
Sample Rate and Delay Time Accuracy | ±200 ppm over any ≥1 ms time interval | |
Delay Time Range | Zero to 50 s |