Agilent 8565EC Spectrum Analyzer

Measurement Performance, Capability, and Quality
The new Agilent Technologies 8560 EC-Series of portable, color display, RF and microwave spectrum analyzers offers the measurement capability and performance previously found only in larger, more expensive benchtop analyzers. The EC-Series analyzers combine superior phase noise performance, 1 Hz resolution bandwidths, synthesized tuning, sensitivity, and wide dynamic range in a package built to withstand diverse environmental conditions.

The Agilent 8560 EC-Series provides a complete solution for adjacent channel power (ACP) testing of burst carrier signals using digital modulation as well as the ability to measure occupied bandwidth percentage. Time-gated signal analysis lets you easily measure a wide variety of time-varying signals, such as pulsed RF, time division multiple access (TDMA), interleaving, and burst modulation.

Features:
  • Color LCD display
  • Fast, digital resolution bandwidths of 1, 3, 10, 30, and 100 Hz
  • Adjacent channel power, channel power, carrier power, occupied bandwidth percentage, and timegated measurements standard
  • Precision timebase and 1 Hz counter resolution
  • MIL-PRF-28800F, class 3 rugged design

Specifications:
Frequency
Frequency Range (Internal Mixing) 9 kHz to 50 GHz
Frequency Range (External Mixing):
18 GHz to 325 GHz in 12 waveguide bands (not available with 8560EC Option 002)
Frequency Reference Accuracy
Temperature Stability
± 1x10
Aging (per year)
± 1x10
Setability
± 1x10
Warmup (nominal), 5 minute ± 1x10– 7 ; 15 minute ± 1x10
Frequency Readout Accuracy (N = L0 Harmonic)) Span > 2 MHz x N: ± (freq. readout x freq. ref. accuracy + 5% x span +15% x RBW +10 Hz) Span ≤ 2 MHz x N: ± (freq. readout x freq. ref. accuracy +1% x span +15% x RBW +10 Hz)
General Specifications
Environmental Military Specs: Class 3 MIL-rugged
Calibration Interval: Two years (8560EC/61EC/62EC/63EC); one year (8564EC/65EC)
Warmup Time: 5 minutes in ambient conditions
Temperature: –10˚ to +55˚ C (operating); –51˚ to +71˚ C (not operating)
Humidity: 95% at 40˚ C for five days
Rain Resistance: Drip-proof at 16 liters/hour/sq. ft.
Altitude: 15,000 ft. (operating); 50,000 ft. (not operating)
Pulse Shock (half sine): 30 g for 11 ms duration
Transit Drop: 8-inch drop on six faces and eight corners
Electromagnetic Compatibility: Conducted and radiated interference in compliance with CISPR Pub. 11 (1990). Meets MIL-STD-461C, part 4, with certain exceptions
Power Requirements 115 Vac Operation: 90 to 140 V rms, 3.2 A rms max., 47 to 440 Hz
230 Vac Operation: 180 to 250 V rms, 1.8 A rms max., 47 to 66 Hz
Maximum Power Dissipation: 180 W (8560EC/61EC/62EC/63EC); 260 W (8564EC/65EC)
Audible Noise
(nominal): < 5.0 Bels power at room temp. (ISO DP7779) Dimensions (w/o handle, feet, cover): 187 mm H x 337 mm W x 461 mm D (7.48 in x 13.34 in x 18.44 in)
Weight
(carrying, nominal): 20 kg (44 lb)