Anritsu MS2070A 3 GHz Field Master Handheld Spectrum Analyzer

Anritsu MS2070A 3 GHz Field Master Handheld Spectrum Analyzer

Anritsu MS2090A 9GHz Field Master Pro Handheld Spectrum Analzyer

Anritsu MS2090A 9GHz Field Master Pro Handheld Spectrum Analzyer

Anritsu MS2080A 4 GHz Field Master Handheld RF Spectrum Analyzer

$15,295.00
Availability: In stock
SKU
6C215

Condition: Refurbished / Calibrated

Warranty: 1 YEAR WARRANTY

Shipping: USA & Worldwide

  • Anritsu Field Master MS2080A Spectrum Analyzer (BC# -L/T*)
  • Stylus
  • AC Adapter/Charger
  • Power Cord
  • USB PC Interface Cable
  • DC Car Charger
  • User Manual On CD-ROM
  • Carrying Case
Request a Quote for

Have questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

We are commited to win your business!

Anritsu MS2080A 4 GHz Field Master Handheld RF Spectrum Analyzer

Anritsu is proud to introduce the next line in handheld spectrum analyzer with real-time spectrum analysis. With frequency coverage up to 6 GHz, the new Field Master™ MS2080A completely redefines the standards for portable handheld analyzers, setting another new industry benchmark for performance and accuracy. The new MS2080A is the culmination of over 60 years of microwave test and measurement equipment development, using the very latest technologies to deliver accuracy and precision in measurements previously reserved only for benchtop instruments.

 

Features:

  • 5GNR FDD/TDD FR1 Analyzer
  • Real-Time Spectrum Analyzer
  • LTE FDD and TDD Analyzer
  • Spectrogram
  • Gated Sweep
  • AM/FM Audio Demodulation
  • Field Strength
  • EIRP
  • Occupied Bandwidth
  • Channel Power
  • Adjacent Channel Power
  • Spectral Emissions Mask
  • Signal Strength and RSSI
  • Carrier Aggregation
  • Coverage Mapping in SPA, 5GNR, and LTE
  • Carrier-to-Interference

 

Specifications:

Smart Measurements Field Strength: Measures field strength (dBm/m 2 , dBW/m2, dBV/m, dBmV/m, dBμV/m, V/m, W/m2, W/cm 2 , A/m) with antenna gain vs. frequency plot
Channel Power: Measures the total power and power spectral density within a specified bandwidth
Occupied Bandwidth: Measures the 99 % to 1 % power channel of a signal
Adjacent Channel Power: Measures the channel power of the adjacent channel
Spectral Emission Mask: Standards based limits for wireless emissions.
Carrier-to-Interference (C/I): Measures the ratio of power (dB) in an RF carrier to the interference power in the channel.
Setup Parameters Frequency: Center/Start/Stop, Frequency Step, Frequency Offset, Gestures
Span: Span (Manual/Increment 1, 2, 5), Full Span, Last Span, Zero Span
Amplitude: Reference Level (Manual/Auto and Offset), Scale/Division, Y-Axis Unit (dBm, dBW, dBV, dBmV, dBμV, dBA, V, W, A), Preamp (On/Off), Attenuation (Auto/Manual), Attenuation Level, Field Strength, Gestures
Bandwidth: RBW/VBW (Auto/Manual), VBW Type (Linear/Logarithmic), RBW:VBW Ratio, SPAN:RBW Ratio
Sweep: Single/Continuous, Restart, Sweep Once, Sweep to N, Auto/Manual Time, Gated Sweep
Spectrogram Number of Lines: 142
Trace Time/Position Cursor: Up to six cursors (display historical trace data by trace position or time)
Cursor State: Active, Hold/View, Blank
Color Setup: Color Scale Top/Bottom Range, Reference Hue, Preset Setup
Trace Functions Traces: Up to Six Traces
Trace Type: Clear/Write, Average (2 to 1000), Max Hold, Min Hold, Rolling Average, Rolling Max Hold, Rolling Min Hold
Trace Math: T1-T2, T2-T1 (when T5 and T6 are selected)
Trace Mode: Active, Hold/View, Blank
Detector Type per Trace: Peak, RMS/Avg, Negative, Sample, Normal
Trace Record: Record live samples with manual tagging to internal or external storage
Trace Playback: Play recorded samples from internal or external storage; set playback interval
CSV Logging: Record live or playback traces in CSV format for post processing
Trigger Functions (Zero span only) Sources: Free Run, Video, External
Settings: Level, Delay, Holdoff, Periodic, Slope (Rising/Falling), Hysteresis
Marker Functions Markers: Up to 12 Markers
Marker Measurements: Amplitude, Frequency (swept spectrum display) Amplitude, Time (Zero Span)
Marker Mode: Normal, Delta, Fixed
Delta Marker: Relative to any Normal or Fixed Marker
Marker Function: None, Noise, Frequency Counter (1 Hz, 100 mHz, 10 mHz, 1 mHz resolutions), Quasi-Peak (per CISPR 16-1-1)
Marker Trace: Assign Marker to any Trace
Peak Search: Peak Search, Next Peak, Next Peak Left, Next Peak Right, Next Point Left, Next Point Right
Peak Search Setup: Peak Threshold, Peak Excursion
Marker: Mkr Center, Mkr Ref Level
Marker Table: Up to 12 Markers Showing Marker, Mode, Function, Trace, Frequency, Amplitude, Delta Frequency & Offset
Limit Line Functions Limit Setup: Upper/Lower, Limit On/Off, Limit Alarm On/Off, Set Default Limit Line, Absolute/Relative
Limit Line Edit: Frequency, Amplitude, Add Point, Add Vertical, Delete Point, Next Point Left/Right
Limit Line Move: Center, X-Offset (Hz), Left, Right, Y-Offset, Up, Down, To Marker 1, Marker 1 Offset (dB)
Limit Line Envelope: Select Envelope (Upper/Lower), Set Envelope, Envelope Points (2-41), Amplitude Offset, Shape (Square/Slope)
Frequency MS2080A-0704: 9 kHz to 4 GHz (Option 704)
MS2080A-0706: 9 kHz to 6 GHz (Option 706)
Tuning Resolution: 1 Hz
Span: 10 Hz to max frequency, Zero Span
Frequency Reference: Internal, GNSS, External
Internal Frequency Reference: Standard TCXO:
Aging: ±1.0 x 10–6 per year
Accuracy: ±2.8 x 10–7 (-10°C ±55°C) plus aging
External Frequency Reference: 10 MHz, -10 dBm to +10 dBm
Bandwidth Analysis Bandwidth: 20 MHz (standard), 40 MHz (Option 102)
Resolution Bandwidth (RBW): 1 Hz to 5 MHz, 10 Hz to 10 MHz in zero span (standard), 10 Hz to 20 MHz in zero span (Option 102)
Video Bandwidth (VBW): 1 Hz to 5 MHz, 10 Hz to 10 MHz in zero span (standard), 10 Hz to 20 MHz in zero span (Option 102)
RBW Selectivity: 4:1 nominal (–60 dB / –3 dB)
CISPR Bandwidth: Resolution bandwidth when using Quasi-Peak marker function: 200 Hz, 9 kHz, and 120 kHz
Sweep Manual Sweep: Maximum sweep time is 3600 s (1 hour)
Sweep Points: 10 to 10,001 (1001 in zero span)
Sweep Rate (non-zero span): 32 GHz/s (standard), 45 GHz/s (Option 102)
Zero Span Sweep Time: 60 ns to 3600 s in zero span
Sweep Time Accuracy: ±2 % in zero span
Spectral Purity – SSB Phase Noise Offset from 1 GHz RF Input
10 kHz: –94 dBc/Hz
100 kHz: –97 dBc/Hz
1 MHz: –123 dBc/Hz
Spurs Residual Spurious: < –110 dBm (RF input terminated, 0 dB input attenuation, > 10 MHz, preamp On)
< –95 dBm (RF input terminated, 0 dB input attenuation, 10 MHz to 4 GHz, preamp Off)
< –90 dBm (RF input terminated, 0 dB input attenuation, 4 GHz to 6 GHz, preamp Off)
Input-Related Spurious: < –70 dBc (0 dB attenuation, –30 dBm input)
Exceptions, typical: < –68 dBc @ 700 MHz to 3300 MHz with 2086 MHz Input
< –65 dBc @ 2*(F1 –1484) MHz, where 3140 MHz < F1 < 3580 MHz
< –68 dBc @ F1 – 2086 MHz where 2100 MHz < F1 < 4970 MHz
Local-Oscillator Related Spurious: < –60 dBc nominal for offsets > 1 MHz
Amplitude Ranges Dynamic Range: 105 dB typical at 1 GHz, 2/3 (TOI-DANL) in 1 Hz RBW
Measurement Range: DANL to +30 dBm
Display Range: 1 to 15 dB/div in 1 dB steps, ten divisions displayed
Reference Level Range: –150 dBm to +30 dBm
Attenuator Resolution: 0 to 50 dB, 5 dB steps
Reference Level Offset: 99.9 dB external loss to 99.9 dB external gain
Maximum Continuous Input: +30 dBm CW, ±50 VDC (≥ 10 dB attenuation)
+23 dBm CW, ±50 VDC (< 10 dB attenuation)
+10 dBm CW, ±50 VDC (preamp ON)
Damage Level: 5 W (+37 dBm) to 6 GHz