Anritsu MS2721B HandHeld Spectrum Master Analyzer w/ OPT: 20
$2,949.00
Availability:
In stock
SKU
9S21
Condition: Refurbished / Calibrated
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- Anritsu MS2721B Spectrum Master (BC# 41193-L/T*)
- Installed Options:
- Option 20: Tracking Generator
- Installed Options:
- Ethernet Cable
- (2) BNC Cables
- USB PC Interface Cable
- AC Adapter/ChargerPower Cord
- User Manual On CD-ROM
- Carrying Case w/ Strap
Request a Quote for
Anritsu MS2721B Spectrum Analyzer
The Spectrum Master MS272xB is a synthesizer-based handheld spectrum analyzer that provides quick and accurate measurement results. Measurements can be easily made by using the main instrument functions: frequency, span, amplitude, and bandwidth. Dedicated keys for common functions and a familiar calculator-type keypad are available for fast data entry.
The Spectrum Master is designed for monitoring, measuring, and analyzing signal environments. Typical measurements include: in-band interference and transmit spectrum analysis, digital television transmitter testing and analysis (in models MS2721B and MS8911B), plus cell site and 802.11a/b/g interference testing. Options are available (for models MS2721B, MS2723B, and MS2724B for RF, advanced demodulation, and over-the-air (OTA) measurement. A full range of marker capabilities (such as peak, center, and delta functions) are provided for faster, more comprehensive analysis of displayed signals. Upper and lower multi-segmented limit lines are available to create quick, simple pass/fail measurements. A menu option provides for an audible alert when a limit value is exceeded.
Continuous frequency coverage from 9 kHz to 7.1 GHz gives the wireless professional the performance needed for the most demanding measurements in harsh RF and physical environments.Whether you need spectrum monitoring, AM and FM broadcast proofing, WiFi and WiFi5 installation and testing, RF and microwave signal measurements or cellular signal measurements, the MS2721B Spectrum Master is the tool to make your job easier and more productive.
Features:
Specifications:
The Spectrum Master MS272xB is a synthesizer-based handheld spectrum analyzer that provides quick and accurate measurement results. Measurements can be easily made by using the main instrument functions: frequency, span, amplitude, and bandwidth. Dedicated keys for common functions and a familiar calculator-type keypad are available for fast data entry.
The Spectrum Master is designed for monitoring, measuring, and analyzing signal environments. Typical measurements include: in-band interference and transmit spectrum analysis, digital television transmitter testing and analysis (in models MS2721B and MS8911B), plus cell site and 802.11a/b/g interference testing. Options are available (for models MS2721B, MS2723B, and MS2724B for RF, advanced demodulation, and over-the-air (OTA) measurement. A full range of marker capabilities (such as peak, center, and delta functions) are provided for faster, more comprehensive analysis of displayed signals. Upper and lower multi-segmented limit lines are available to create quick, simple pass/fail measurements. A menu option provides for an audible alert when a limit value is exceeded.
Continuous frequency coverage from 9 kHz to 7.1 GHz gives the wireless professional the performance needed for the most demanding measurements in harsh RF and physical environments.Whether you need spectrum monitoring, AM and FM broadcast proofing, WiFi and WiFi5 installation and testing, RF and microwave signal measurements or cellular signal measurements, the MS2721B Spectrum Master is the tool to make your job easier and more productive.
Features:
- 9 kHz to 7.1 GHz Input
- 1 Hz to 3 MHz RBW Range
- Very Low Phase Noise (ヨ100 dBc/Hz Maximum at 10 kHz offset at 100 kHz to 7.1 GHz)
- Built-in AM/FM/SSB Demodulator
- Built-in Preamplifier
- 65 dB Step Attenuator
- Input Protected to 20 Watts
- True RMS Detection
- 2+ Hours of Battery Life
- 3.1 kg (<6.9 lbs)
- 3G Modulation options
- Tracking Generator option
- GPS Receiver option
- Quasi-peak detector and CISPR bandwidths
Specifications:
Frequency | |
Frequency Range: | 9 kHz to 7.1 GHz |
Tuning Resolution: | 1 Hz |
Frequency Reference: | |
Aging: | ±1 ppm per 10 years |
Accuracy: | ±0.3 ppm (25º C ± 25º C) + aging |
Frequency Span: | 10 Hz to 7.1 GHz plus 0 Hz (zero span) |
Span Accuracy: | Same as frequency reference accuracy |
Sweep Time: | |
Zero span: | 10 ms to 600s |
Spans >0 Hz: | Sweep time is automatically optimized. Can be manually increased |
Sweep Time Accuracy: | ±2% in zero span |
Sweep Trigger: | Free run, Single, Video, External |
Resolution Bandwidth: | (ヨ3 dB) 1 Hz to 3 MHz in 1-3 sequence ±10%, 200 Hz, 9 kHz, 120 kHz when quasi-peak detector selected, 10MHz demodulation bandwidth |
Video Bandwidth: | : (ヨ3 dB) 1 Hz to 3 MHz in 1-3 sequence |
SSB Phase Noise: | |
Offset from carrier 10, 20 and 30 kHz |
Max ヨ100 dBc/Hz |
100 kHz | ヨ102 dBc/Hz |
Amplitude | |
Measurement Range: | DANL to +30 dBm |
Display Range: | 1 to 15 dB/div in 1 dB steps. Ten divisions displayed |
Amplitude Units: | |
Log Scale Modes: | dBm, dBV, dBmv, dBµV |
Linear Scale Modes: | nV, µV, mV, V, kV, nW, µW, mW, W, kW |
Attenuator Range: | 0 to 65 dB |
Attenuator Resolution: | 5 dB steps |
Absolute Amplitude Accuracy: | |
Power levels: | =ヨ50 dBm, =35 dB input attenuation |
9 kHz to =10 MHz | ±1.5 dB |
>10 MHz to 4 GHz | ±1.25 dB |
>4 to 7.1 GHz | ±1.75 dB |
40 to 55 dB input attenuation | |
9 kHz to =10 MHz | ±1.5 dB |
>10 MHz to 4 GHz | ±1.75 dB |
>4 to 6.5 GHz | ±1.75 dB |
>6.5 to 7.1 GHz | ±2 dB |
60 to 65 dB input attenuation | |
9 kHz to =10 MHz | ±1.5 dB |
>10 MHz to 6.5 GHz | ±1.75 dB |
>6.5 to 7.1 GHz | ±3 dB |
Preamplifier on, 0 or 10 dB input attenuation | |
100 kHz to 4 GHz | ±1.5 dB |
>4 to 7.1 GHz | ±1.75 dB |
Second Harmonic Distortion (0 dB input attenuation, ヨ30 dBm input): |
|
0.05 to 1.4 GHz | ヨ50 dBc |
>1.4 to 2 GHz | ヨ70 dBc |
>2 GHz | ヨ80 dBc |
Third Order Intercept (TOI): (ヨ20 dBm tones 100 kHz apart, ヨ20 dBm Ref level, 0 dB input attenuation, preamplifier off |
|
Frequency | Min |
600 MHz | +7 dBm |
3.5 GHz | +9 dBm |
Frequency | Typical |
50 MHz to 300 MHz | >8 dBm |
>300 MHz to 2.2 GHz | 10 dBm |
>2.2 to 2.8 GHz | >15 dBm |
>2.8 to 4.0 GHz | >10 dBm |
>4.0 to 7.1 GHz | >13 dBm |
Dynamic Range 2/3 (TOI-DANL) in 1 Hz RBW | |
600 MHz | 95 dB min |
3.5 GHz | 96 dB min |
General | |
RF Input VSWR: | (=10 dB input attenuation) 2.0:1 max, 1.5:1 typical |
Maximum Continuous Input: | (=10 dB input attenuation) +30 dBm |
Input Damage Level | =10 dB input attenuation, >+43 dBm, ±50 Vdc <10 dB input attenuation, >+23 dBm, ±50 Vdc2 Input protection relay opens at >30 dBm with =10 dB input attenuation and at approximately 10 to 23 dBm with <10 dB input attenuation. with limited dV/dt. |
ESD Damage Level: | (=10 dB input attenuation) >10 kV |
External Reference Frequencies: | 1, 1.2288, 1.544, 2.048, 2.4576, 4.8, 4.9152, 5, 9.8304, 10, 13 and 19.6608 MHz at ヨ10 dBm to +10 dBm |
Battery Life: | 2.5 hours typical |
Display | Bright daylight-viewable color transmissive LCD: Full SVGA, 8 in |
Languages | Built-in English, Spanish, Italian, French, German, Japanese, Korean, and Chinese. The instrument also has the capability to have two customized languages installed from Master Software Tools. |
Marker Modes | 6 Markers, 9 Modes: Normal, Delta, Marker to Peak, Marker to Center, Marker to Reference Level, Next Peak Left, Next Peak Right, All Markers Off, Noise Marker, Frequency Counter Marker (1 Hz resolution), Markers Tracking or Fixed, Marker 1 reference for all deltas. |
Sweeps | Full span, Zero span, Span Up/Span Down |
Detection | Peak, Negative peak, Sample, RMS, Quasi-peak |
Memory | Trace and Setup storage is limited only by the capacity of the installed Compact Flash card or USB Flash drive. For a 512 MB card, storage is greater than 13000 spectrum analyzer traces and over 10000 setups |
Traces | Displayed Traces: Three Traces with trace overlay. Trace A is always the live data; Traces B and C can be either stored data or traces which have been mathematically manipulated. Also Trace C can show max hold or min hold. |
Interfaces | Type N female RF connector for Spectrum Analyzer input Type N female RF connector for optional Tracking Generator Reverse polarity BNC jack for optional GPS antenna connector BNC female connectors for ext. reference and ext. trigger 5-pin Mini-B USB 2.0 for data transfer to a PC USB 2.0 Host connector used with PSN50 High Accuracy Power Meter and USB Flash Drives RJ45 connector for Ethernet 10/100 Base T 2.5 mm 3-wire headset connector |
Size and Weight | |
Size: | 313W x 211H x 77D mm (12W x 8H x 3D in.) |
Weight: | 3.1 kg (<6.9 lbs.) typica |
Environmental | |
Operating: | ヨ10º C to 55º C, humidity 85% or less |
Storage: | ヨ51º C to 71º C |
Altitude: | 4600 meters, operating and non-operating |
For full Anritsu Spectrum Master MS2721B product specifications, please click here: Anritsu Spectrum Master MS2721B