BK Precision 2658A Handheld Spectrum Analyzer
$5,995.00
Availability:
In stock
SKU
9S34
Condition: Refurbished / Calibrated
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- BK Precision 2658A Spectrum Analyzer (BC# 23856-L/T*)
- Armored Test Port Cable
- Battery Pack
- AC Adapter/Charger
- Power Cord
- USB PC Interface Cable
- User Manual On CD-ROM
- Carrying Case (G)
Request a Quote for
BK Precision 2658A Spectrum Analyzer
The 2650A series of compact, light weight and cost-effective spectrum analyzers are for quick and precise signal investigations, especially away from the bench. At a weight of only 4 pounds (1.8 kg), these instruments are by far the lightest full-featured handheld spectrum analyzers available, yet they deliver performance and features comparable to a full-size bench spectrum analyzers and can operate up to 4 hours on a single battery.
With their ease of use, measurement flexibility and unmatched portability, the 2650A series are an indispensible tool for engineers and technicians who conduct field measurements in the 50 kHz to 8.5 GHz range.
For full BK Precision 2658A specifications, please click here: BK Precision 2658A
The 2650A series of compact, light weight and cost-effective spectrum analyzers are for quick and precise signal investigations, especially away from the bench. At a weight of only 4 pounds (1.8 kg), these instruments are by far the lightest full-featured handheld spectrum analyzers available, yet they deliver performance and features comparable to a full-size bench spectrum analyzers and can operate up to 4 hours on a single battery.
With their ease of use, measurement flexibility and unmatched portability, the 2650A series are an indispensible tool for engineers and technicians who conduct field measurements in the 50 kHz to 8.5 GHz range.
- Impressive 4 hour battery life; easy-to-replace rechargeable Lithium-Ion battery
- Fast sweep speed, minimum 10 ms, to help locate and identify elusive, transient interference signals
- DANL (displayed average noise level) of -127 dBm
- Large easy-to-read color display, 640 x 480 resolution
- Measurement functions: channel/adjacent channel power, occupied bandwidth, electric and magnetic field strength
- Convenient Auto tune function
- Conveniently store measurement results and screen shots in bitmap format to USB flash drive
- Installation, maintenance, and troubleshooting of wireless communication systems such as W-CDMA/CDMA, GSM, WLAN, WiMAX and Bluetooth
- Detection of signal interference and undesired emissions
- TV and broadcasting
- Antenna alignment
Frequency section | |
Center frequency | |
Frequency range | 50 kHz to 8.5 GHz |
Setting | 20 kHz |
resolution | Settable with rotary encoder, numeric or function key |
Accuracy | within ± (30+20T) kHz±1 dot @frequency span: 200 kHz to 10 MHz, RBW 3 kHz, 23±5 °C within ± (60+300T) kHz±1dot @frequency span: 20 MHz to max. frequency (3.3 GHz or 8.5 GHz), RBW 100 kHz, 23±5 °C T: sweep time (s) |
RBW frequency error | within ±4 kHz @ 3 kHz, 10 kHz, 30 kHz within 20% of RBW @ RBW: 100 kHz, 300 kHz within 10% of RBW @ RBW: 1 MHz, 3 MHz |
Frequency span | |
Range | 0 Hz (zero span), 200 kHz to 5 GHz (1-2-5 step) and 8.5 GHz (Full span) |
Accuracy | within ±3% ±1dot @ sweep time of 0.3s, 23±5 °C |
Display resolution # of points per trace | 501dots on LCD screen, 1001dots readout via PC (501 dots are visible on the display, 1001 dots of trace data are captured internally and can be transferred to a PC via USB device interface.) |
Bandwidth | 3 dB bandwidth |
Range | 3 kHz to 3 MHz (1-3 sequence) and AUTO |
Accuracy | ±20% |
Selectivity | 1:12 (typical, 3 dB : 60 dB) |
Video bandwidth | 100 Hz to 1 MHz (1-3 step) and AUTO |
SSB phase noise | -90 dBc/Hz (typical) @100 kHz offset, RBW : 3 kHz, VBW : 100 Hz, sweep time: 1 s |
Spurious response | Less than -60 dBc |
Harmonics | less than -40 dBc @ ≥ 100 MHz |
Amplitude section | |
Reference level | |
Range | +10 to -60 dBm (1dB step) |
Accuracy | within ± 0.8 dB ± 1 dot @ center frequency : 100 MHz, RBW : 3 MHz, VBW : 1 MHz, REF : -15 dBm, 23±5 °C |
Unit | dBm, dBV, dBmV, dB V, dB V/m, dB A/m (dB V/m and dB A/m are used for measurement functions) |
Average noise level | -127 dBm (typical) @ CF : 1 GHz, RBW : 3 kHz, VBW : 100 Hz, Ref. level < - 40 dBm ( preamp automatically ON) |
Frequency response | within ±2.0 dB ± 1 dot @50 kHz to 100 MHz within ±1.0 dB ± 1 dot @100 MHz to max. frequency (3.3 GHz or 8.5 GHz) |
Input impedance | 50 ? |
Input VSWR | < 2.0 |
Input attenuator | |
Operating range | 0 to 25 dB (1 dB step), coupled with reference level |
Switching error | ±0.6 dB @100 MHz |
RBW switching error | ±0.6 dB |
Display resolution (vertical) | 381 dots/10 div |
Scale | 10 dB/div, 5 dB/div, 2 dB/div |
Accuracy | ±(0.2 dB+1dot)/2 dB ±(0.4 dB+1dot)/5 dB ±(0.8 dB+1dot)/10 dB ±(1.8 dB+1dot)/83 dB |
Input damage level | +27 dBm (CW average power), 25 VDC |
Sweep section | |
Sweep time | |
Range (1-3 step) | 10 ms to 30 s and AUTO @frequency span : 0 to 2 GHz 30 ms to 30 s and AUTO @frequency span : 5 GHz, full span |
Accuracy | Within ±0.1 % ±1 dot @frequency span: 0 to 5 GHz within ±2.5 % ±1 dot @ full span |
Trigger | |
Trigger mode | AUTO (Available only for zero span) |
Trigger source | Internal and External |
Input voltage range | 1 to 10 Vp-p |
Frequency range | DC to 5 MHz |
Input coupling | DC coupling |
Trigger level | approx. 0.56 V(fix) |
Input RC | approx. 10 k? / less than 15 pF |
Input damage level | ±50 V(DC+AC peak) |
Input connector | SMA(J) |
Detection mode | Positive peak, Negative peak, Sample |
Function | |
Marker measurement | NORM displays frequency (8 digits max) and level (4 digits max) at marker point. DELTA displays frequency difference and level difference between two markers. |
Peak search | NORM: searches peak point within 10div (full freq. range). This mode also supports NEXT peak ( up to 10). ZONE: searches peak point within a zone designated by center and width |
Calculations | NORM, MAX HOLD, MIN HOLD, AVERAGE, OVER WRITE MAX/MIN HOLD: 2 to 1024 AVERAGE: 2 to 256 |
Measurements | Measure Channel power, Adjacent channel leakage power, Occupied frequency bandwidth, Electric field strength (requires antenna), Magnetic field strength (requires magnetic field probe) |
Auto tuning | Automatically scans the full bandwidth, sets center frequency to the maximum signal within full span and centers it onscreen. Automatically adjusts reference level, RBW, VBW and sweep time to optimum values |
Save/Load | |
Save | Saves 200 reference spectrum measurements and 200 setups |
Load | Loads one reference spectrum measurement and one setup |
General | |
RF Input connector | N(J) connector |
USB Communication | |
Protocol | USB version 1.1 (will also work with USB 2.0 devices) |
Device Connector | B plug |
Transfer rate | 12 Mbps |
Hard copy | USB printer (option) connected to A plug (host) enables hard copy of screen. |
Host connector | A plug |
Display | |
Display | 5.7 inch color TFT LCD |
Backlight | LED backlight |
Number of dots | 640(H) x 480(V) dots |
Power supply | |
Lithium-ion battery | 7.4 V/5000 mAh Li-Ion battery (MB400) |
Charge function | Capable of charging during power-off. Indicates 4 conditions with two color LED |
Charge indicator | 5 level indicator of remaining battery charge |
Environmental and Size | |
Operating temperature | 0 to 50° C (guaranteed at 23±10 °C, without soft carrying case) |
Operating humidity | less than 40 °C/80 %RH (guaranteed at less than 33 °C/70 %RH, without soft carrying case) |
Storage temperature | -20 to 60 °C, less than 60 °C/70 %RH |
Dimensions | 6.38 (W) x 2.80 (H) x 10.43 (D) inch, 162(W) x 71(H) x 265(D) mm (excluding projections, protection bumper and stand) |
Weight | approx. 1.8 kg or 4 lbs (including battery) |
For full BK Precision 2658A specifications, please click here: BK Precision 2658A