Fluke RPM90

Fluke RPM90 Inductive Pick-Up / External Trigger

Fluke 189 II kit with accessories

Fluke 189 II True RMS Digital Multimeter

Fluke 787 ProcessMeter DMM Loop Calibrator

$695.00
Availability: In stock
SKU
10H09

Condition: Refurbished / Calibrated

Warranty: 1 YEAR WARRANTY

Shipping: USA & Worldwide

  • Fluke 787 ProcessMeter (BC# 43821-L/T*)
  • (2) 1000V Hard Point Test Leads
  • User Manual On CD-ROM
  • Carrying Case (G)
Request a Quote for

Have questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

We are commited to win your business!

 Fluke 787 Process Meter
The Fluke-787 ProcessMeter combines a DMM and a Loop Calibrator in one rugged, handheld tool - for about what you would expect to pay for a loop calibrator alone. The 787 starts with the Fluke-87 DMM you know and trust, then adds ability to measure, source, and simulate DC loop current with 0.05% accuracy and 1 micro-amp resolution. When you carry the ProcessMeter, you're ready for any loop troubleshooting or calibration task. And since itメs dust- and splash-resistant and EMI-shielded, the 787 can survive in your tool belt.

Features:
  • DMM and Loop Calibrator in one tool
  • Precision 1000V, 440 mA digital Multimeter
  • Measure AC and DC volts, AC and DC current, resistance, continuity, and frequency 0.1% DC voltage accuracy 0.05% DC current accuracy, 1 uA resolution to 30 mA Simultaneous mA and % of scale readout True RMS AC voltage measurement to 1 kHz Frequency measurement to 20 kHz Min/Max/Average/Hold/Relative modes Diode Test and Continuity Beeper
  • DC current source / loop calibrator 20 mA current source / simulator Simultaneous mA and % of scale readout Manual Step (25%, Coarse, Fine) plus Auto Step and Auto Ramp
  • Clear LCD with backlight; 4,000 counts (30,000 counts for DC current)
  • Externally accessible battery for easy battery changes
  • V overload protection on V, ohms, frequency
  • V overload protection on mA, backed-up by 440 mA 1000V fuse
  • Designed to meet 1000 volt IEC 1010 Cat III standards
 
Measurement Function Range and Resolution Best Accuracy (% of reading + LSD)
VDC 400.0 mV, 4.000V, 40.00 V, 400.0 V, 1000 V 0.1% + 1
VAC (true RMS) 400.0 mV, 4.000V, 40.00 V, 400.0 V, 1000 V  0.7% + 4
mADC 30.000 mA .05% + 2
A DC 1.000 A (0.440A continuous)  0.2% + 2
A AC 1.000 A (0.440A continuous) 1% + 2
Resistance 400.0 Ohms, 4.000k, 40.00 k, 400.0 k, 4.0 M, 40M  0.2% + 1
Frequency
(0.5 Hz to 20 kHz)
199.99 Hz, 1999.9 Hz, 19.999 kHz .005% + 1
Diode Test 2.400 V (shows diode voltage drop)  2% + 1
Continuity  Beeps for resistance < approx 100 ohms
 
Output Function Range and Resolution Drive Capability Accuracy (% of span)
DC Current Output

(Internal battery operation)
0.000 to 20.000 mA or 4.000 to 20.000 mA (selectable at power-up)

Over-range to 24.000 mA
12 V compliance, or,

500 Ohms, @ 20 mA
.05%
DC Current Simulate
(Ext. 24 Volt loop supply)
0.000 to 20.000 mA or 4.000 to 20.000 mA, (selectable at power-up)
Over-range to 24.000 mA
1000 Ohms, @ 20 mA 05%
Current adjustment modes Manual: Coarse, Fine, 25% step
Automatic: Slow Ramp, Fast Ramp, 25% step


General Specifications:
Maximum voltage applied between any jack and Earth Ground: 1000 V RMS
Storage Temperature: -40°C to 60°C
Operating Temperature: -20°C to 55°C
Temperature Coefficient: 0.05 x (specified accuracy) per °C (for temperatures < 18°C or > 28°C)
Relative Humidity: 95% up to 30°C; 75% up to 40°C; 45% up to 50°C; 35% up to 55°C
Vibration: Random, 2g, 5-500 Hz.
Shock: 1 meter drop test
Safety: Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No.1010..1-92 Over-voltage Category III
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L)
Size, with holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
Weight: 369 g (13 oz); with holster and Flex-Stand 638 g (22.5 oz)
Battery: Single 9V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
Battery Life: 50 hours typical (measurement), 12 hours typical (sourcing 12 mA)



For full Fluke 787 specifications, please click here: Fluke 787