GE KrautKramer USM 32 X B Ultrasonic Flaw Detector
$3,295.00
Availability:
Out of stock
SKU
2K14
Condition: Refurbished / Calibrated
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
Request a Quote for
- KrautKramer USM 32X B Ultrasonic Flaw Detector (BC# 32157-L/T*)
- AC Adapter/Charger
- Power Cord
- RS232 PC Interface Cable
- Hard Copy User Manual
- Case For Unit w/ Strap
- Carrying Case (G)
KrautKramer USM 32X Ultrasonic Flaw Detector
The USM 32X is a lightweight and compact ultrasonic flaw detector especially suitable for locating and evaluating material defects & measuring wall thicknesses.
The different instrument versions
The USM 32X is available in several versions which are desgined for different applications:
Features:
Specifications:
For full KrautKramer USM 32X product specifications, please click here: KrautKramer USM 32X
The USM 32X is a lightweight and compact ultrasonic flaw detector especially suitable for locating and evaluating material defects & measuring wall thicknesses.
The different instrument versions
The USM 32X is available in several versions which are desgined for different applications:
- Basic version USM 32X B
- Standard version, for universal ultrasonic test tasks.
- Version USM 32X L
- Special version, designed for ultrasonic testing of large components (up to 10 m / 390") and for the frequency range from 0.1 to 18 MHz. Typical applications using this version are thickness measurement and flaw detection on: large forgings, cast blocks (steel, aluminium), workpieces made of non-ferrous metals, plastics, composites (bonding test)
- USM 32X F
- Special version, designed for the measurement of fat and muscle thicknesses in stockbreeding.
Features:
- low weight (2.2 kg including lithium-ion battery) and compact size
- long operating time (> 10 hours) by means of lithium-ion battery with internal and external charging possibility
- handy – equipped with handle/non-slip, ratcheting prop-up stand
- rotary knobs for direct adjustment of gain as well as for changing the currently selected function
- two independant gates for accurate wall thickness measurements from the workpiece surface up to the first echo, or between two backwall echoes, including measurement on coated workpieces with a resolution of 0.01 mm (up to 100 mm), referred to steel (0.1 mm in USM 32X F)
- 5.7", 1/4 VGA-TFT color display to display the digitized signals (320 × 240 pixels, 115 × 86 mm)
- display of sound path and amplitude for gate A plus 1 reading zoomed in the A-scan, user-configurable
- data memory: 100 data sets, documentation possibility via a printer or in file
Specifications:
Size (HxWxD) | 176 x 255 x 105 mm |
Weight | 2,2 kg |
Storage temperature | -20 - +60 °C |
Temperate range for mains operation | 0 - 40 °C |
Temperature range for battery operation | 0 - 60 °C |
Warm-up period | 10 minutes at 25° C |
Voltage range for battery operation | V= 8,6 5,4 13 8 |
Possible power supplies | Battery, mains (external power supply) |
Voltage range with power supply | V= 9,5 13 |
Power consumption | VA 5,5 - 8 |
Low battery warning | yes |
Environmental Protection | IP 54 |
Shock resistivity | Shock resistance according to DIN IEC 68, 6 ms, 60g shocks per axis |
Vibration | Vibration resistance according to DIN IEC 68, 1-150 Hz, 2g, 20 cycles per axis |
Technical specifications according to EN 12668-1 | |
Damping | 342Ý 45Ý |
Capacity | Lower Power: 220 High Power: 1000 |
Pulse Repetition Frequency | Min: 20% Average: 4-1000 Hz Max: 20% |
PRF Mode | Manual, automatically linked to range in 10 steps |
PRF step | >1.5m 4 Hz <1.5m 100 Hz |
Operating Modes | Single-, dual mode |
Characteristics of Transmitter Pulse | |
Pulse voltage | Min: -376 Average: -342 Max: -307 1000 Hz DAMPING = low POWER = high |
Pulse rise time | Min: 10, 38 ns Average: 12,21 ns Max: 14, 04ns |
Pulse duration | Min: 229 ns Average: 254,5 ns Max: 280 ns |
Effective output impedance | Min: 16 Average: 20 Max: 24 |
Pulse Voltage: | Min: -266 Average: -242 Max: -217 1000 Hz DAMPING = high POWER = high |
Pulse rise time | Min: 13,71 Average: 16,13 Max: 18,55 |
Pulse duration | Min: 151 Average: 167,5 Max: 184 |
Effective output impedance | Min: 19,2 Average: 24 Max: 28,8 |
Pulse voltage | Min: -342 Average: -311 Max: -280 1000 Hz DAMPING = low POWER = low |
Pulse rise time | Min: 9,07 Average: 137, 5 Max: 12,27 |
Effective output impedance | Min: 17,6 Average: 22 Max: 26, 4 |
For full KrautKramer USM 32X product specifications, please click here: KrautKramer USM 32X