GE USM GO Ultrasonic Flaw Detector

GE USM GO Ultrasonic Flaw Detector

GE DMS GO Ultrasonic Thickness Gauge

GE DMS GO Ultrasonic Thickness Gauge

GE KrautKramer USM 32 X B Ultrasonic Flaw Detector

$3,295.00
Availability: Out of stock
SKU
2K14

Condition: Refurbished / Calibrated

Warranty: 1 YEAR WARRANTY

Shipping: USA & Worldwide

Request a Quote for

Have questions or doubts? Submit the following request and expect our detailed quote within 24 hours!

We are commited to win your business!

  • KrautKramer USM 32X B Ultrasonic Flaw Detector (BC# 32157-L/T*)
  • AC Adapter/Charger
  • Power Cord
  • RS232 PC Interface Cable
  • Hard Copy User Manual
  • Case For Unit w/ Strap
  • Carrying Case (G)
KrautKramer USM 32X Ultrasonic Flaw Detector
The USM 32X is a lightweight and compact ultrasonic flaw detector especially suitable for locating and evaluating material defects & measuring wall thicknesses.

The different instrument versions
The USM 32X is available in several versions which are desgined for different applications:
  • Basic version USM 32X B
    • Standard version, for universal ultrasonic test tasks.
  • Version USM 32X L
    • Special version, designed for ultrasonic testing of large components (up to 10 m / 390") and for the frequency range from 0.1 to 18 MHz. Typical applications using this version are thickness measurement and flaw detection on: large forgings, cast blocks (steel, aluminium), workpieces made of non-ferrous metals, plastics, composites (bonding test)
  • USM 32X F
    • Special version, designed for the measurement of fat and muscle thicknesses in stockbreeding.

Features:
  • low weight (2.2 kg including lithium-ion battery) and compact size
  • long operating time (> 10 hours) by means of lithium-ion battery with internal and external charging possibility
  • handy – equipped with handle/non-slip, ratcheting prop-up stand
  • rotary knobs for direct adjustment of gain as well as for changing the currently selected function
  • two independant gates for accurate wall thickness measurements from the workpiece surface up to the first echo, or between two backwall echoes, including measurement on coated workpieces with a resolution of 0.01 mm (up to 100 mm), referred to steel (0.1 mm in USM 32X F)
  • 5.7", 1/4 VGA-TFT color display to display the digitized signals (320 × 240 pixels, 115 × 86 mm)
  • display of sound path and amplitude for gate A plus 1 reading zoomed in the A-scan, user-configurable
  • data memory: 100 data sets, documentation possibility via a printer or in file

Specifications:
Size (HxWxD) 176 x 255 x 105 mm
Weight 2,2 kg
Storage temperature -20 -  +60 °C
Temperate range for mains operation 0 - 40 °C
Temperature range for battery operation 0 - 60 °C
Warm-up period 10 minutes at 25° C
Voltage range for battery operation V=  8,6 5,4 13 8
Possible power supplies Battery, mains (external power supply)
Voltage range with power supply V= 9,5 13
Power consumption VA 5,5 - 8
Low battery warning yes
Environmental Protection IP 54
Shock resistivity Shock resistance according to DIN IEC 68, 6 ms, 60g shocks per axis
Vibration Vibration resistance according to DIN IEC 68, 1-150 Hz, 2g, 20 cycles per axis
Technical specifications according to EN 12668-1
Damping 342Ý 45Ý
Capacity Lower Power: 220
High Power: 1000
Pulse Repetition Frequency Min: 20%
Average: 4-1000 Hz
Max: 20%
PRF Mode Manual, automatically linked to range in 10 steps
PRF step >1.5m 4 Hz
<1.5m 100 Hz
Operating Modes Single-, dual mode
Characteristics of Transmitter Pulse
Pulse voltage Min: -376
Average: -342
Max: -307
1000 Hz DAMPING = low POWER = high
Pulse rise time Min: 10, 38 ns
Average: 12,21 ns
Max: 14, 04ns
Pulse duration Min: 229 ns
Average: 254,5 ns
Max: 280 ns
Effective output impedance Min: 16
Average: 20
Max: 24
Pulse Voltage: Min: -266
Average: -242
Max: -217
1000 Hz DAMPING = high POWER = high
Pulse rise time Min: 13,71
Average: 16,13
Max: 18,55
Pulse duration Min: 151
Average: 167,5
Max: 184
Effective output impedance Min: 19,2
Average: 24
Max: 28,8
Pulse voltage Min: -342
Average: -311
Max: -280
1000 Hz DAMPING = low POWER = low
Pulse rise time Min: 9,07
Average: 137, 5
Max: 12,27
Effective output impedance Min: 17,6
Average: 22
Max: 26, 4



For full KrautKramer USM 32X product specifications, please click here: KrautKramer USM 32X