Fluke Networks OFTM-5631 SM Fiber OTDR Module

Fluke Networks OFTM-5631 SM Fiber OTDR Module

Fluke Networks OFTM-5611 MM Fiber OTDR Module

Fluke Networks OFTM-5611 MM Fiber OTDR Module

JDSU Viavi T-BERD 4118RMA65 SM Fiber OTDR 1650nm

$1,895.00
Availability: In stock
SKU
3L111

Condition: Refurbished / Calibrated

Warranty: 1 YEAR WARRANTY

JDSU 4118RMA65 Specs


The MA OTDR module meets the challenges of commissioning a complete metro ring, troubleshooting a bend in a distribution frame, or qualifying high-port-count optical splitters in passive optical networks (PON). Its impressive technical specifications combined with a wide range of test functions give technicians with the best solution to more efficiently deploy or repair fiber links in the field.

The MA module’s optical performance combined with the complete T-BERD/MTS platform suite ensures testing is done right—the first time.

Standard test features include:

  • Automatic macrobend detection
  • Summary results table with pass/fail analysis
  • Bidirectional OTDR analysis
  • FastReport onboard report generation

Features & Benefits

  • Up to 40 dB dynamic range
  • PON-optimized to test up to 1x32 splitter ratio, and up to 1x64 splitter ratio with the FTTH-SLM application
  • Integrated CW light source and power meter
  • FiberComplete™ compatible
  • Ready for SLM, FTTA-SLM, and FTTH-SLM intelligent optical application software
  • Instantly detects traffic upon connecting live fiber


Specifications:

General (Typical at 25°C)
Weight 0.35 kg (0.77 lb)
Optical Interfaces
Interchangeable optical connectors FC, SC, DIN, LC, and ST
Technical Characteristics
Laser safety class (21 CFR) Class 1
Distance units Kilometers, feet, and miles
Group index range 1.30000 to 1.70000, in 0.00001 steps
Number of data points Up to 128,000 data points
Distance measurement Automatic or dual cursor
Display range 0.5 to 260 km
Cursor resolution 1 cm
Sampling resolution 4 cm
Accuracy ±1 m ±sampling resolution ±1.10-5 x distance (Excluding group index uncertainties)
Attenuation Measurement
Automatic, manual, 2-point, 5-point, and LSA
Display range 1.25 to 55 dB
Display resolution 0.001 dB
Cursor resolution 0.001 dB
Linearity ±0.03 dB/dB
Threshold 0.01 to 5.99 dB in 0.01 dB steps
Reflectance/ORL Measurements
Reflectance accuracy ±2 dB
Display resolution 0.01 dB
Threshold 11 to −99 dB in 1 dB steps
MA OTDR Module (Typical at 25°C)
Central wavelength 1650±20 nm
Pulse width 3 ns to 20 µs
RMS dynamic range 37 dB
Event dead zone 90 cm
Attenuation dead zone 4m



For full JDSU 4118RMA65 Fiber OTDR Module product specifications, please click here: 4118RMA65

JDSU 4118RMA65 SM 1650nm Fiber OTDR Module

The JDSU 4118RMA65 OTDR metro access module can be used with the T-BERD MTS 2000, 4000 & 5800 Platfoms
(Platforms NOT INCLUDED!)

Dynamic Range:
1650 nm : 37 dB

JDSU Viavi T-BERD 4118RMA65 SM Fiber OTDR JDSU 4118 RMA65 OTDR Module Models On Sale Warranty Calibration Backed by The Best Service and Lowest Prices in the Industry.

 

 