MIE PDR-1000AN Personal DataRAM Airborne Particulate Monitor
Condition: Refurbished
Warranty: 1 YEAR WARRANTY
Shipping: USA & Worldwide
- MIE PDR-1000AN Personal DataRAM (BC# 9021A-L/T*)
- Battery Charger
- User Manual On CD-ROM
- Carrying Case
MIE PDR-1000 AN Personal DataRAM Monitor
The Thermo Scientific personal DataRAM pDR-1000AN Monitor is a passive sampling, light-scattering nephelometer that measures real-time mass concentrations of dust, smoke, mists, and fumes in industrial as well as other indoor and outdoor environments.
Nephelometry, a light-scattering technology, is highly sensitive and incorporates a pulsed, high output, nearinfrared light emitting diode source, as well as a silicon detector/hybrid preamplifier, collimating optics and a source reference feedback PIN silicon detector.
The pDR-1000AN monitor features a high measurement range of 0.001 to 400 mg/m3 (auto-ranging), a 400,000-fold span. An optical feedback and stabilized sensing system, sets the standard for sensitivity, long-term stability and reliability.
Simple zeroing, with particle-free air, is quick using the included zeroing kit. The internal firmware controls an automatic calibration check. A gravimetric calibration can be performed by weighing a separate filter sample and programming in the calibration constant to provide the same reading.
The pDR-1000AN monitor is an ultracompact, durable and self-contained instrument designed for hand-held or beltworn unattended operation. There are no moving parts which provides for silent, dependable operation.
Selectable alarm levels with built-in audible signal and switched output, a RS-232 communications port, and a programmable analog concentration output are all part of this versatile instrument.
Key Features:
• Measure airborne particulate concentration in real-time
• Highest measurement range of any real-time personal particulate monitor
• Simple zeroing and calibration
• No moving parts, silent operation
• Compact, durable, and self-contained
Specifications | |
Concentration Measurement Range | 0.001 to 400 mg/m3 range (auto ranging)1 |
Scattered Coefficient Range | 1.5 x 10-6 to 0.6m-1 (approx.) @ lambda= 880nm (not displayed) |
Precision/Repeatability Over 30 Days | ±0.5 of reading or ± 0.015 mg/m3 , whichever is larger, for 1 second averaging time |
±0.5 of reading or ± 0.015 mg/m3 , whichever is larger for 10 second averaging time | |
±0.2% of reading or ± 0.005 mg/m3 , whichever is larger for 60 second averaging time | |
Accuracy1 | ±5% of reading ± precision (traceable to SAE Fine Test Dust) |
Resolution | 0.1% of reading or 0.001 mg/m3 , whichever is larger |
Particle Size Range of Max. Response | 0.1 to 10μm |
Aerodynamic Particle Cut-Point Range | 10μm nominal |
Concentration Display Updating Interval | 1 second |
Concentration Display Averaging Time2 | 1 to 60 seconds (user selectable) |
Data Logging Averaging Periods3 | 1 second to 1 hour |
Total # of Data Points Logged in Memory | 13,391 |
Number of Data Tags | 99 (maximum) |
Logged Data | Average concentration, time/date, and data point number |
Readout Display | LCD 16 characters (4 mm height) x 2 lines |
Serial Interface | RS-232 / 4,800 baud |
Computer Requirements | IBM-PC compatible, 486 or higher, Windows 95® or higher, ≥ 8 MB memory |
Hard disc drive 3.5" floppy, VGA or higher resolution monitor | |
Analog Output | 0 - 5V and 4-20mA |
Physical Dimensions | 6" (153mm) H x 3.6" (92mm) W x 2.5" (63mm) D |
Weight | 18oz. (0.5kg) |